Agilent Technologies B1500A Manual de servicio Pagina 97

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 110
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 96
Module 6
Low Current Measurement
6-25
z You will notice that sub 10 fA measurements are possible
z Set the jumper leads for the bipolar transistor
z Insert the device and perform a zero cancel offset
z Get the GUMMEL setup in the Demo preset group
To Get Started:
Class Exercise
Low Current Gummel Plot
The following pages will lead you through the setup and measurement procedure for a low current
Gummel measurement. Single triaxial cable connections to the force lines of SMUs 2, 3, and 4 are
fine for this test. Kelvin triaxial cables are not necessary.
Vista de pagina 96
1 2 ... 92 93 94 95 96 97 98 99 100 101 102 ... 109 110

Comentarios a estos manuales

Sin comentarios