Detector Constant. The primary considerations for selecting a calibration
method are measurement accuracy and equipment availability.
Technique
Use the current Detector Constant
Derive From Double Sided Spur
Derive From Single Sided Spur
Equipment
Required
None
Source with calibrated AM or
calibrated amplitude
modulator.
Directional coupler, RF
analyzer, RF source.
Accuracy
Very Good
Very Good
* Accuracy depends on which technique was used to determine the Detector Constant in
memory.
Use the Current Detector Constant
Introduction
This calibration method configures the HP 3048A to use the Detector
Constant currently in memory. This eliminates the time spent calibrating
the AM Detector Constant. The AM Detector Constant in memory is the
value that was determined by the last AM Noise measurement made
(unless a stored Test File has been loaded). The Detector Constant value
appears in the Calibration Process display.
The Detector Constant is valid for repeat measurements of similar devices
as long as the same detector is being used and the input power levels here
remained the same (within ±0.3 dB).
2-90 Calibr Process: AM Noise