NOTE
When measuring a Two Port device, the AM Noise con-
tribution of the Source can affect the measurement. It is
recommended that a Source with Low AM Noise be used
to ensure good results.
Phase Shifter: When the source frequency cannot be adjusted, a Phase
Shifter is required for achieving quadrature for the Residual Two Port
measurement.
DUT:
The DUT blocks allow for either one or two devices to be configured
in the measurement paths. The DUT blocks are positioned by pressing the
softkey. You must configure a device in both measurement paths
Place DUTs
when measuring devices that translate frequency or add large amounts of
phase delay. The physical connection of where the DUT is placed with
respect to the Phase Shifter is not important (unless the Two Port device
to be measured translates the frequency). In this case, the frequency range
of the Phase Shifter needs to be considered.
DUT placement is done only to help visualize the measurement config-
uration and specify a correct Connect Diagram including DUT's. DUT
placement has no affect on any facet of system operation.
Quadrature: The System allows you to specify how quadrature (90° phase
shift) will be set. Pressing the Quadr Method toggles the quadrature selection
between adjustment of the source frequency and the manual method. (If
the source is under System Control and entered in the System Control
Table, the frequency adjustment will be automatic.) The manual method
requires a phase shifter (or adjustable line stretcher) to achieve quadrature.
2-106 Source Control: $N w/o PLL