Agilent Technologies InfiniiVision 7000B Series Guía de usuario Pagina 212

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 428
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 211
6 Measurements and Math Functions
212 InfiniiVision Oscilloscopes User’s Guide
Overshoot
Overshoot is distortion that follows a major edge transition expressed as a percentage of
Amplitude. The X cursors show which edge is being measured (edge closest to the trigger
reference point).
A
UTOMATIC OVERSHOOT MEASUREMENT
Rising edge overshoot =
local Maximum - D Top
Amplitude
----------------------------------------------------------
X 100
Falling edge overshoot =
Base - D local Minimum
Amplitude
-----------------------------------------------------------
X 100
local Maximum
Overshoot
Overshoot
local Minimum
Top
Base
Vista de pagina 211
1 2 ... 207 208 209 210 211 212 213 214 215 216 217 ... 427 428

Comentarios a estos manuales

Sin comentarios