Agilent Technologies InfiniiVision 7000B Series Guía de usuario Pagina 348

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 428
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 347
11 Mask Test
348 InfiniiVision Oscilloscopes User’s Guide
Building a Mask File
The following mask uses all eight regions. The most difficult part of creating a mask file is
normalizing the X and Y values from the time and voltage values. This example shows an
easy way to convert voltage and time to normalized X and Y values in the mask file.
The following mask file produced the mask shown above:
MASK_FILE_548XX
"All Regions"
/* Region Number */ 1
/* Number of vertices */ 4
-12.50, MAX
-10.00, 1.750
10.00, 1.750
12.50, MAX
/* Region Number */ 2
/* Number of vertices */ 5
-10.00, 1.000
-12.50, 0.500
-15.00, 0.500
-15.00, 1.500
Vista de pagina 347
1 2 ... 343 344 345 346 347 348 349 350 351 352 353 ... 427 428