10 Serial Decode/Lister
308 InfiniiVision Oscilloscopes User’s Guide
Interpreting I
2
C Decode
Angled waveforms show an active bus (inside a packet/frame).
Mid-level blue lines show an idle bus.
In the decoded hexadecimal data:
Address values appear at the start of a frame.
Write addresses appear in light-blue along with the “W” character.
Read addresses appear in yellow along with the “R” character.
Restart addresses appear in green along with the “S” character.
Data values appear in white.
A” indicates Ack (low), “~A indicates No Ack (high).
Decoded text is truncated at the end of the associated frame when there is
insufficient space within frame boundaries.
Red dots in the decode line indicate that more data can be displayed. Scroll or expand
the horizontal scale to view the data.