6 SCPI Command Reference
136 N9340A User’s Guide
Amplitude Step
:TGENerator:AMPLitude:STEP <ampt>
:TGENerator:AMPLitude:STEP?
This command sets the amplitude step size of the
tracking generator output. This command is valid
after tracking generator is enabled.
Amplitude Offset
:TGENerator:AMPLitude:OFFSet <ampt>
This command sets the amplitude offset of the
tracking generator output. This command is valid
after tracking generator is enabled.
*RST State: 1 dB
Range: 1 to 10 dB
Min. Increment 1 dB
Available unit: dB
Key Access: {Amptd Step}
*RST State: 0.00 dB
Range: –88.40 to 327.60 dB
Min. Increment 0.01 dB
Available unit: dB
Key Access: {Amptd Offset}