Making Measurements 3
N9340A User’s Guide 29
Use the knob or the arrow keys to further reduce
the resolution bandwidth and better resolve the
signals.
As you decrease the resolution bandwidth, you
improve resolution of the individual signals and
also increased the sweep timed. For fastest
measurement times, use the widest possible
resolution bandwidth.
Under factory preset conditions, the resolution
bandwidth is coupled to the span.
Figure 4 Resolving signals of equal amplitude