Channel Commands
Command
:CHANnel<number>:OFFSet[?] <offset_value>
For TDR and TDT applications, you may change the offset to magnify offset
in some circumstances. This command is used to set the magnify offset as
well as the offset.
<number>
An integer, 1-4, indicating the slot in which the channel resides.
<offset_value>
Offset value at center screen; can be volts, amperes, or other units.
Command
:CHANnel<number>:RANGe[?] <range_value>
For TDR and TDT applications, you may change the range to magnify range
in some circumstances. This command is used to set the magnify range as
well as the range.
<number>
An integer, 1-4, indicating the slot in which the channel resides.
<range_value>
Full-scale voltage of the specified channel number.
Command
:CHANnel<number>:SCALe[?] <scale_value>
For TDR and TDT applications, you may change the scale to magnify scale in
some circumstances. This command is used to set the magnify scale as well
as the scale.
<number>
An integer, 1-4, indicating the slot in which the channel resides.
<scale_value>
Vertical scale of the channel in units per division.