85108L System Manual
Operation
Pulsed-RF Measurements Overview
3-17
Figure 3-13 Amplifier Gain, Frequency (point-in pulse)
The dynamic range can be increased using IF averaging, but, given the system noise floor with the
wide IF bandwidth, an averaging factor of about 256 averages is the maximum value that should be
used for most measurements. Using an averaging factor greater than 256 will not result in any
increase in visible dynamic range.