Agilent Technologies 83624A Manual de servicio Pagina 87

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 122
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 86
85108L System Manual
System Service and Troubleshooting
Troubleshooting Strategy
6-5
Front Panel Keys Method
Set the rear panel switches on the source as shown in Figure 6-1.
Figure 6-1 8360 Series Source Rear Panel Switches
8340/41 Source GPIB Address Check
These sources are set by front panel keys. The procedure sets the source GPIB address to 19 (the
address of RF source 1).
Procedure
1. Turn on the source.
2. Press
SHIFT LOCAL 1 9 HZ.
3. Turn the source off.
Vista de pagina 86