Count Accuracy Troubleshooting
Agilent Technologies
PSA Series Spectrum Analyzers
Test and Adjustment Software
Count Accuracy Troubleshooting
Related Adjustments
10 MHz Internal Frequency Reference Adjustment
If a failure occurs, be sure the 10 MHz reference is connected.
Failure to meet this specification can be due to a faulty A11 Reference assembly if the
adjustment does not solve the problem.
http://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/Troubleshooting/troublecount.htm [5/29/2003 8:47:42 AM]