Agilent Technologies 83630A Especificaciones Pagina 118

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 133
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 117
Third Order Intermodulation Distortion Troubleshooting
Agilent Technologies
PSA Series Spectrum Analyzers
Test and Adjustment Software
Third Order Intermodulation Distortion
Troubleshooting
Related Adjustments
Lowband Mixer Bias Adjustment
Switched LO Distribution Amplifier Adjustment
Second LO Power Adjustment
If TOI fails to meet specifications, suspect the first mixer of the band which fails (A20
Low Band or A19 RYTHM). Next suspect the A10 Third Converter assembly.