Agilent Technologies
PSA Series Spectrum Analyzers
Test and Adjustment Software
50 MHz Calibrator Amplitude Adjustment
An RF substitution method is made using the 50 MHz, -25 dBm calibrator Z5602A. The
level of the calibrator is read by the PSA and then the PSA’s internal 50 MHz calibrator is
measured. The difference is stored in memory as a correction.
Required Test Equipment
Test Equipment Model Number
E4440A
E4443A, E4445A
E4446A
E4448A
50 MHz Calibrator
Z5602A 0ption 51
(Opt H35 for BAB
Opt H24 for E4446A, E4448A)
X X
50 MHz Calibrator Amplitude Adjustment Setup