304 Chapter7
Replacing Major Assemblies
A2 Display
11.Slowly remove the display assembly from the front of the
instrument. Continue to route W2 and W3 through the slot (4) as
the display is removed.
Figure 7-2 A2 Display Replacement
12.Remove the screws and the fiber washer (6) that secure the A2
Display to its enclosure (7). See Figure 7-4.
13.Slowly pull the display out of its enclosure while routing W2 and W3
through the opening in the enclosure (8).