Chapter 2 61
Making Adjustments
6. Cal Attenuator Error
2 dB Check
13.Press the following analyzer keys.
AMPLITUDE, 16, dBm.
14.Set the amplitude of the synthesizer/level generator to 17 dBm.
15.Press the following analyzer keys.
PEAK SEARCH
MARKER DELTA, MARKER DELTA
MKR FCTN, MK TRACK ON OFF (ON)
16.MKR amplitude reading should be 0.0 ±0.01. If it is not, repeat
step 15.
17.Set the analyzer by pressing
AMPLITUDE, 14, dBm.
18.Set the amplitude of the synthesizer/level generator to 15 dBm.
19.Press
MKR on the analyzer.
20.Record the MKR amplitude reading in Table 2-2. This is the 2 dB
attenuator step error of the A12 amplitude control assembly.
4 and 8 dB Step Error
21.Repeat step 13 to step 20 for attenuator steps 4 dB and 8 dB. Use
Table 2-2 for synthesizer values and analyzer reference-level values.
Table 2-2 A12 Amplitude Control Assembly Attenuator Errors
SynthesizerSettings(dBm) Reference Level Settings
(dBm)
Attenuator
Step (dB)
Error
Attenuator
OFF
Attenuator
ON
Attenuator
OFF
Attenuator
ON
19 18 18 17 1 ____________
17 15 16 14 2 ____________
18 14 17 13 4 ____________
19 11 18 10 8 ____________