Agilent Technologies 8590 L-Series Especificaciones Pagina 620

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 674
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 619
620 Chapter14
Analyzer Messages
Interpreting Analyzer Messages
Before removing the A16 processor/video assembly, check all assemblies
that are identified in Table 14-2.
1. Turn the instrument power off and remove each assembly, one at a
time.
2. Turn the instrument on.
If the ADC 2 V fault remains, the instrument preset at power-up
displays the error message.
If the error message goes away, the last assembly removed is the
cause of the problem.
If the error message remains after the removal of all related
assemblies, suspect the A16 processor/video assembly.
Check the ADC 2 V reference voltage on the A16 processor/video
assembly. Refer to 8590 Series Analyzers Component-Level Repair
Service Guide for the location of the 2 V reference circuit in the
analog-to-digital function block.
A 2 V reference failure can also be checked manually, using the service
diagnostics. Press
CAL, More 1 of 4, More 2 of 4, Service Diag, More 1,
2v REF DETECTOR.
Vista de pagina 619
1 2 ... 615 616 617 618 619 620 621 622 623 624 625 ... 673 674

Comentarios a estos manuales

Sin comentarios