Agilent Technologies N5980A Manual de usuario Pagina 7

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 17
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 6
J-BERT N4903A High-Performance Serial BERT Data Sheet 7
Pattern generator key characteristics:
Differential outputs for data, clocks and trigger
Variable output voltages covering LVDS, ECL, CML
Transitions times < 20 ps
Clean pulses with Jitter < 9 ps pp
High-precision delay control input to inject jitter from
an external source
Calibrated and integrated jitter injection
(option J10, J20, both retrofitable)
Subrate clocks for generating any reference clock
Pattern sequencing and 32 Mbit pattern
SSC clocks for computer buses (option J11)
Figure 12: Clean output signal. 10 Gb/s, LVDS levels.
Figure 13: Pattern generator setup screen with graphical dis-
play of signal levels.
Table 1: Output characteristics for J-BERT N4903A generator.
All timing parameters are measured at ECL levels
Range of operation 150 Mb/s to 12.5 Gb/s (opt.
C13) Can be programmed up to
13.5 Gb/s.
150 Mb/s to 7 Gb/s (opt. C07)
< 620 MHz only with external
clock.
Frequency accuracy ± 15 ppm typical
Format NRZ, normal or inverted
Amplitude/Resolution 0.10 V to 1.8 V, 5 mV steps
Adresses LVDS, CML, PECL,
ECL (terminated to 1.3 V/
0 V/-2 V), low voltage CMOS
Output voltage window -- 2.0 V to +3.0 V
Predefined levels ECL, PECL (3.3 V), LVDS, CML
Transition times
(20% to 80%) < 20 ps
(10% to 90%)
2)
< 25 ps
Jitter 9 ps pp typical with disabled jitter
sources
Clock/data delay range ± 0.75 ns in 100 fs steps
External termination -- 2 V to +3 V
voltage
3)
Crossing point Adjustable 20% to 80% typical to
emulate duty cycle distortions
Single error inject Adds single errors on demand
Fixed error inject Fixed error ratios of 1 error in
10
n
bits, n = 3, 4, 5, 6, 7, 8, 9
Interface
1)
Differential or single-ended,
DC coupled, 50
Connector 2.4 mm female
1)
Unused outputs must be terminated with 50to GND.
2)
At 10 Gb/s and 7 Gb/s
3)
For positive termination voltage or termination to GND, exter-
nal termination voltage must be less than 3 V below VOH. For
negative termination voltage, external termination voltage must
be less than 2 V below VOH. External termination voltage must
be less than 3 V above VOL.
Figure 11: Generator connectors on front panel.