16
dataTec ▪ Ferdinand-Lassalle-Str. 52 ▪ 72770 Reutlingen ▪ Tel. 07121 / 51 50 50 ▪ Fax 07121 / 51 50 10 ▪ [email protected] ▪ www.datatec.de
1616
Up to 110 MHz (4-Terminal Pair) SMD
Electrode dimensions
Short compensationOpen compensation Inserting a DUT
Compensation and Measurement: Open and short compensations
are recommended before measurement. Short compensation is
performed by bringing down the shorting plate (which is already on
the fixture) to short all 4 terminals. Open compensation is
performed by bringing down both the open plate and the shorting
plate to separate the high terminals from the low terminals. After
performing open and short compensations, the DUT is inserted into
the test fixture. The figures below show how compensation and
measurement are performed.
Diameter of probe pin is 0.19 mm
Diameter of Blade electrode’s contact spot is 0.5 mm
Spacing between Blade electrode and probe pin is 0.11 mm
SMD can be between 1.6 and 8.0 mm in width