Agilent Technologies DC110 Especificaciones Pagina 52

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 80
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 51
48
dataTec ▪ Ferdinand-Lassalle-Str. 52 ▪ 72770 Reutlingen ▪ Tel. 07121 / 51 50 50 ▪ Fax 07121 / 51 50 10 ▪ [email protected] ▪ www.datatec.de
4848
compensation function of the measurement instrument first. Next,
open compensation is performed by not placing anything in the
device insertion hole. Short compensation is performed by placing
the furnished shorting device in the device insertion hole. After per-
forming open and short compensations in combination with the elec-
trical length compensation, the DUT is inserted into the device inser-
tion hole. Once the measurement of the DUT is complete, remove the
DUT from the fixture. The following figures show how compensation
and measurement is performed.
Up to 3 GHz (7 mm) SMD
Compensation and measurement
Vista de pagina 51
1 2 ... 47 48 49 50 51 52 53 54 55 56 57 ... 79 80

Comentarios a estos manuales

Sin comentarios