16453A Dielectric Material Test
Fixture
Terminal Connector: 7 mm
Dimensions (approx.): 130(H) x 50(W) x 60(D) [mm]
Weight (approx.): 600 g
Measurement Accuracy (including the E4991A):
Up to 3 GHz (7 mm) Material
Description: The 16453A is designed for accurate dielectric constant
and loss tangent measurements on the E4991A. It employs the par-
allel plate method, which sandwiches the material between two
electrodes to form a capacitor. E4991A measures the capacitance
created from the fixture, and option E4991A-002 firmware calculates
the relative complex permittivity. Adjustment to insure parallel elec-
trodes is required when using the 16451B. This adjustment is not
required with 16453A because the fixture has a flexible electrode
that adjusts automatically to the material surface.
Applicable Instruments:E4991Awithpt.E4991A-0024291B*
withpt.4291B-002***
*denotestheinstrumentisobsolete.
**applicableinalimitedreuencyrange.
Frequency: 1 MHz to 1 GHz
Maximum Voltage: ±42 V peak max. (AC+DC)
Operating Temperature: -55°C to 200°C
When Option E4991A-007 temperature characteristic test kit is used
with E4991A, the operating temperature range is between -55°C and
+150°C.
Material Size:
Furnished Accessories:
Compensation and Measurement: Open, short and load compensa-
tions are recommended before measurement. Open compensation is
performed by separating the high and the low electrodes from each
other. Short compensation is performed by connecting the high and
low electrodes together. Load compensation is performed by using
the furnished load material. After performing open, short and load
compensations, the material under test is inserted into the test fix-
ture.
Typical Permittivity (er’) Measurement Accuracy
(@ thickness = 1 mm)
Typical Loss Tangent (tan ) Measurement Accuracy
(@ thickness = 1 mm)
E4991A with 16453A
Diameter ≥ 15 mm
Thickness 0.3 mm ~ 3 mm
Description P/N Qty.
Fixture Holder 16453-01213 1
Load 16453-60021 1
Tweezers 8710-2081 1
Carrying Case 16453-60011 1
Operation and Service Manual 16453-90010 1