Agilent Technologies E1326B Manual de usuario Pagina 68

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 291
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 67
Additional Measurement Functions
The following MEASure and CONFigure statements can be substituted into
the example programs to make measurements other than DC voltage.
Additional Stand-Alone
Multimeter Functions
The following statements can be substituted into the program “Making a
Single Measurement” on page 42.
!AC voltage.
OUTPUT 70903;"MEAS:VOLT:AC?"
!
4-wire resistance.
OUTPUT 70903;"MEAS:FRES?"
!
4-wire thermistor (type = 2252, 5000, 10000).
OUTPUT 70903;"MEAS:TEMP? FTH,type"
!
4-wire RTD (type = 85, 92).
OUTPUT 70903;"MEAS:TEMP? FRTD,type"
The following statements can be substituted into the programs where the
faceplate terminals are used to make multiple reading bursts or multiple
burst measurements.
!AC voltage.
OUTPUT 70903;"CONF:VOLT:AC"
!
4-wire resistance.
OUTPUT 70903;"CONF:FRES"
!
4-wire thermistor (type = 2252, 5000, 10000).
OUTPUT 70903;"CONF:TEMP FTH,type"
!
4-wire RTD (type = 85, 92).
OUTPUT 70903;"CONF:TEMP FRTD,type"
Additional Scanning
Multimeter Functions
The following statements can be substituted into the program “Scanning a
Channel List” on page 46.
!AC voltage.
OUTPUT 70903;"MEAS:VOLT:AC? (@channel_list)"
!
2-wire resistance.
OUTPUT 70903;"MEAS:RES? (@channel_list)"
!
4-wire resistance (channels available are 00 through 07).
OUTPUT 70903;"MEAS:FRES? (@channel_list)"
!
Thermocouple (type = B, E, J, K, N14, N28, R, S, T).
OUTPUT 70903;"MEAS:TEMP? TC,type,(@channel_list)"
Chapter 3 Using the HP E1326B/E1411B Multimeter 69
Vista de pagina 67
1 2 ... 63 64 65 66 67 68 69 70 71 72 73 ... 290 291

Comentarios a estos manuales

Sin comentarios