13
Measuring Preamble and DL
Instructions Keystrokes
On the P-Series:
1. Measure Preamble at 104 µs
duration (See Figure 11).
{Trace Control}, {X Start –200 µs & X
Scale 300 µs}. {Gate Control}. Press [
] to allocate marker 2 and [ ]
to allocate marker 1.
2. Measure DL with only Burst 1 is
ON (See Figure 12).
Press [ ] to allocate marker 2 and
[ ] to allocate marker 1.
Figure 11 Preamble duration at 104 µs and its average power
measured at 0.04 dBm which is close to 0 dBm of
input power (see Signal Generator setup on Page 10).
Figure 12 DL average power measures –4.52 dBm at 1.25 ms
where Burst 1 is ON only. Referring to Figure 11, the
Preamble (P) has a 3 dB boost above the DL or nominal
level of the rest of the signal. At times, user’s bursts
can be boosted above 0 dB as well.
Comentarios a estos manuales