Agilent Technologies Agilent E5250A Manual de usuario Pagina 9

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 12
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 8
Altitude
Operating: 0 to 2,000 m (6,561 ft)
Storage: 0 to 4,600 m (15,091 ft)
Power requirement
90V to 264V, 47 to 63 Hz
Maximum VA
4155C and 4156C: 450VA
41501B: 350 VA
Regulatory Compliance
EMC: EN 61326-1:+A1, AS/NZS 2064.1
Safety:
CSA C22.2 NO.1010.1 (1992),
IEC 61010-1:+A2/EN 61010-1:+A2
UL3111-1:1994
Certification:
CE, CSA, NRTL/C, C-Tick
Dimensions
4155C and 4156C: 235mm H ´
426mm W ´ 600mm D
41501B: 190mm H ´
426mm W ´ 600mm D
Weight (approx.)
4155C and 4156C: 21kg
41501B: 16kg (option 412,
HPSMU + 2 ´ PGU)
4155C and 4156C
Furnished Accessories
Triaxial cable, 4 ea. (4155C)
Kelvin triaxial cable, 4 ea. (4156C)
Coaxial cable, 4 ea.
Interlock cable, 1 ea.
Keyboard, 1 ea.
User manual, 1 set
Sample application program disk, 1 ea.
Sample VEE program disk, 1 ea.
VXIplug&play drivers disk for the
4155C and 4156C, 1 ea.
VXIplug&play drivers disk for the
E5250A, 1 ea.
LAN Interface Test Adapter, 1 ea.
Accessory
Specifications
Specification Condition
The “supplemental information” and
“typical”entries in the following specifi-
cations are not warranted, but provide
useful information about the functions
and performance of the instruments
(23°C ±5°, 50% RH).
16440A SMU/Pulse Generator
Selector
The 16440A switches either an SMU
or PGU to the associated output port.
You can expand to 4 channels by
adding an additional 16440A. The
channel 1 PGU port provides a “PGU
OPEN” function, which can disconnect
the PGU by opening a semiconductor
relay. The 16440A cannot work without
two pulse generator units of the
41501A/B (SMU and Pulse Generator
Expander).
Channel configurations:
Two channels (CH1, CH2)
CH1: INPUT ports: 2
(SMU and PGU, PGU port has
additional series semiconductor relay)
OUTPUT port: 1
CH2: INPUT ports: 2 (SMU and PGU)
OUTPUT port: 1
Voltage and Current Range
Input port Max. V Max. I
SMU 200V 1.0A
PGU 40V 0.2A (AC Peak)
Supplemental Information
(at 23°C ± 5°C, 50%RH)
SMU port leakage current:
< 100fA @100V
SMU port residual resistance (typical):
0.2W
SMU port stray capacitance (typical
@1MHz):
Force « Common: 0.3pF
Force « Guard: 15pF
Guard « Common: 130pF
PGU port residual resistance: 3.4W
PGU port OFF capacitance (typical): 5pF
PGU port OPEN capacitance (typical):
700pF (@ 1MHz, Vin - Vout = 0V)
PGU port signal transfer characteristics
Overshoot: < 5% of pulse amplitude
(@20ns leading and trailing time, 50W
pulse generator source impedance,
50pF and 1MW in parallel load).
General Specifications
Dimensions:
50mm H ´ 250mm W ´ 275mm D
Approximate weight: 1.1kg
16441A R-BOX
The 16441A R-BOX adds a selectable
series resistor to the SMU output. You
can select the resistor from the setup
page, and the voltage drop due to the
series resistor is automaticallycompen-
sated for in the measurement result.
Measurement limitations with the
4155C and 4156C and R-BOX:
If you measure device characteristics
including negative resistance over
1MW with the 4155C/4156C and
R-BOX, there is a possibility that
they cannot be measured. There is a
possibility that the 4155C and 4156C
cannot perform measurements
because of DUT oscillations even with
the R-BOX. Whether oscillation occurs
or not depends upon the DUT and
measurement conditions.
Number of SMU channels that can add
a resistor: 2
Resistor values:
1MW, 100kW, 10kW, 0W (each
channel)
Resistance accuracy:
0.3% (at 23°C ±5°C, between input-
output terminal)
Maximum voltage: 200V
Maximum current: 1A (0W selected)
Kelvin connection: Kelvin connection
is effective only when 0W is selected.
Supplemental Information
(at 23°C ± 5°C, 50%RH)
Leakage current: <100fA @ 100V
General Specifications
Dimensions:
72mm H ´ 250mm W ´ 270mm D
Approximate weight: 1.6kg
16442A Test Fixture
Channel Information
SMU:
6 channels (1 triaxial connector
per channel)
3 channels (1 Kelvin triaxial connector
per channel)
VSU:
2 channels (1 BNC connector per
channel)
VMU:
2 channels (1 BNC connector per
channel)
PGU:
2 channels (1 BNC connector per
channel)
GNDU:
1 channel (1 triaxial connector)
INTLK: 6-pin connector
Supplemental Information
(at 23°C ± 5°C, 50% RH)
SMU channel:
Leakage current: 10pA max @200V
(Force or Sense « Common)
Stray capacitance: 15pF max
(Force or Sense « Common)
Stray capacitance: 3pF typical
(Force or Sense « Other SMU)
Residual resistance: 60mW typical
(Force, Sense)
Guard capacitance: 70pF max
(Force or Sense « Guard)
VSU channel residual resistance:
60mW typical
VMU channel residual resistance:
60mW typical
PGU channel characteristic impedance:
50mW typical
GNDU channel residual resistance:
40mW typical (Force, Sense)
General Specifications
Temperature range:
Operating: +5°C to +40°C
Storage: -40°C to +70°C
Humidity range;
Operating: 5% to 80% RH
(no condensation)
Storage: 5% to 90% RH at 65°C
(no condensation)
Dimensions:
140 mm H ´ 260 mm W ´ 260 mm D
Weight (approx.): 2.5kg
9
Vista de pagina 8
1 2 3 4 5 6 7 8 9 10 11 12

Comentarios a estos manuales

Sin comentarios