Agilent Technologies
Agilent 1100 Series
Valves
User’s Guide
Vista de pagina 0
1 2 3 4 5 6 ... 77 78

Indice de contenidos

Pagina 1 - Agilent 1100 Series

Agilent TechnologiesAgilent 1100 Series Valves User’s Guide

Pagina 2

4 Agilent 1100 Series Valves1 Installation and ConfigurationUnpacking the 1100 Series ValveThe M4 micro valve fittings support the Agilent PEEK coated

Pagina 3 - In This Guide…

Agilent 1100 Series Valves 5Installation and Configuration 1Hardware InstallationHardware InstallationThe Agilent 1100 Series External Valves can be i

Pagina 4

6 Agilent 1100 Series Valves1 Installation and ConfigurationHardware InstallationIn order to install a valve on the G1383A Column Organizer the Organi

Pagina 5 - Contents

Agilent 1100 Series Valves 7Installation and Configuration 1Hardware InstallationSetting up the power connection to other Agilent 1100 Series ModulesT

Pagina 6

8 Agilent 1100 Series Valves1 Installation and ConfigurationSoftware ConfigurationSoftware ConfigurationThe 1100 Series Valves can be controlled by th

Pagina 7

Agilent 1100 Series Valves 9Installation and Configuration 1Software ConfigurationFigure 3 Valve configuration

Pagina 8 - Site Requirements

10 Agilent 1100 Series Valves1 Installation and ConfigurationSoftware Configuration

Pagina 9 - Delivery Checklist

11Agilent 1100 Agilent 1100 Series ValvesUser’s GuideAgilent Technologies2OperationOperating the Valve using Agilent ChemStation 12Handheld Controller

Pagina 10

12 Agilent 1100 Series Valves2OperationOperating the Valve using Agilent ChemStationOperating the Valve using Agilent ChemStationAfter the Agilent 110

Pagina 11 - Hardware Installation

Agilent 1100 Series Valves 13Operation 2Operating the Valve using Agilent ChemStationAfter the run is finished, the valve always switches to the start

Pagina 12 - Setting up the CAN connection

Agilent 1100 Series ValvesNotices© Agilent Technologies, Inc. 2002, 2004No part of this manual may be reproduced in any form or by any means (includin

Pagina 13

14 Agilent 1100 Series Valves2OperationOperating the Valve using Agilent ChemStationPosition DescriptionsDefine the Position Description that is used

Pagina 14 - Software Configuration

Agilent 1100 Series Valves 15Operation 2Operating the Valve using Agilent ChemStationNext position after runIf Next position after run is checked, the

Pagina 15 - Figure 3 Valve configuration

16 Agilent 1100 Series Valves2OperationHandheld Controller functionsHandheld Controller functionsWith firmware revision B.03.11 or higher the followin

Pagina 16 - 10 Agilent 1100 Series Valves

Agilent 1100 Series Valves 17Operation 2Handheld Controller functionsSetting the EMF limit and switch counter1 From the startup screen of the Agilent

Pagina 17 - Operation

18 Agilent 1100 Series Valves2OperationHandheld Controller functions

Pagina 18 - Position

19Agilent 1100 Agilent 1100 Series ValvesUser’s GuideAgilent Technologies3Maintenance, Repair, and TroubleshootingStatus Indicators 20Maintenance and

Pagina 19 - Valve Name

20 Agilent 1100 Series Valves3 Maintenance, Repair, and TroubleshootingStatus IndicatorsStatus IndicatorsThe instrument status indicator indicates one

Pagina 20 - Time Table

Agilent 1100 Series Valves 21Maintenance, Repair, and Troubleshooting 3Maintenance and RepairMaintenance and RepairThe maintenance of the Agilent 1100

Pagina 21 - Display instrument actuals

22 Agilent 1100 Series Valves3 Maintenance, Repair, and TroubleshootingMaintenance and RepairReplacing the stator face and the rotor seal for standard

Pagina 22 - Handheld Controller functions

Agilent 1100 Series Valves 23Maintenance, Repair, and Troubleshooting 3Maintenance and Repair2 Align the Stator Ring (4) that the two short pins on th

Pagina 23

Agilent 1100 Series Valves iiiIn This Guide…The Agilent 1100 Series valves provide the user a comprehensive solutions for more flexibility through sol

Pagina 24 - 2Operation

24 Agilent 1100 Series Valves3 Maintenance, Repair, and TroubleshootingMaintenance and RepairReplacing the rotor seal for micro valves (G1162A and G11

Pagina 25 - Troubleshooting

Agilent 1100 Series Valves 25Maintenance, Repair, and Troubleshooting 3Maintenance and RepairReassembling the valve head1 Mount the new Rotor Seal (5)

Pagina 26 - Status Indicators

26 Agilent 1100 Series Valves3 Maintenance, Repair, and TroubleshootingMaintenance and Repair• Stator heads Stator head for G1157 (2 position/10 port

Pagina 27 - Maintenance and Repair

Agilent 1100 Series Valves 27Maintenance, Repair, and Troubleshooting 3Maintenance and RepairEarly Maintenance Feedback (EMF)Depending on the applicat

Pagina 28 - G1158A, G1159A, G1160A)

28 Agilent 1100 Series Valves3 Maintenance, Repair, and TroubleshootingMaintenance and RepairResetting the EMF counter1 Switch to the Diagnostics View

Pagina 29

Agilent 1100 Series Valves 29Maintenance, Repair, and Troubleshooting 3System Errors and TroubleshootingSystem Errors and TroubleshootingSystem Errors

Pagina 30 - Disassembling the valve head

30 Agilent 1100 Series Valves3 Maintenance, Repair, and TroubleshootingSystem Errors and TroubleshootingCAN failureError Cause If the red Pow. fail st

Pagina 31 - Reassembling the valve head

Agilent 1100 Series Valves 31Maintenance, Repair, and Troubleshooting 3System Errors and TroubleshootingUnknown PositionError Cause An unknown positio

Pagina 32 - 26 Agilent 1100 Series Valves

32 Agilent 1100 Series Valves3 Maintenance, Repair, and TroubleshootingSystem Errors and TroubleshootingValve SynchronizationIf the valve failed to sw

Pagina 33 - Setting the EMF limit

33Agilent 1100 Agilent 1100 Series ValvesUser’s GuideAgilent Technologies4Valve ApplicationsAlternating Column Regeneration 34Sample enrichment and Sa

Pagina 34 - Resetting the EMF counter

iv Agilent 1100 Series ValvesChapter 3, “Maintenance, Repair, and Troubleshooting,” starting on page 19 will describe recommended maintenance and repa

Pagina 35 - System Errors

34 Agilent 1100 Series Valves4 Valve ApplicationsAlternating Column RegenerationAlternating Column RegenerationAlternating column regeneration is a co

Pagina 36 - Switch failure

Agilent 1100 Series Valves 35Valve Applications 4Alternating Column RegenerationIf the valve is switched to position 1 the eluent pump delivers the mo

Pagina 37 - Unknown Position

36 Agilent 1100 Series Valves4 Valve ApplicationsAlternating Column RegenerationCapillary Kit (G1156-68711) for alternating column regeneration with G

Pagina 38 - Valve Synchronization

Agilent 1100 Series Valves 37Valve Applications 4Alternating Column RegenerationCapillary Kit (G1156-68713) for alternating column regeneration with G

Pagina 39 - Valve Applications

38 Agilent 1100 Series Valves4 Valve ApplicationsSample enrichment and Sample strippingSample enrichment and Sample strippingSample enrichment and sam

Pagina 40

Agilent 1100 Series Valves 39Valve Applications 4Sample enrichment and Sample strippingFor the sample enrichment phase the valve is switched to positi

Pagina 41

40 Agilent 1100 Series Valves4 Valve ApplicationsSample enrichment and Sample strippingFigure 22 on page 39 illustrates how the Agilent 1100 Series 2

Pagina 42 - (analytical scale)

Agilent 1100 Series Valves 41Valve Applications 4Sample enrichment and Sample strippingTabl e 12 Fittings, screws, and ferrulesDescription Qty part nu

Pagina 43 - (preparative scale)

42 Agilent 1100 Series Valves4 Valve ApplicationsColumn SelectionColumn SelectionWith the Agilent 1100 Series 6 position selection valve (G1159A) and

Pagina 44 - Sample enrichment

Agilent 1100 Series Valves 43Valve Applications 4Column SelectionCapillary Kit (G1156-68712) for column selection with G1159A (analytical scale)Tabl e

Pagina 45 - Sample stripping

contentsAgilent 1100 Series Valves iContents1 Installation and ConfigurationSite Requirements 2Unpacking the 1100 Series Valve 3Damaged Packaging 3Del

Pagina 46

44 Agilent 1100 Series Valves4 Valve ApplicationsSolvent SelectionSolvent SelectionThe 1100 Series 12 position / 13 port valve can be used for solvent

Pagina 47 - Valve Applications 4

Agilent 1100 Series Valves 45Valve Applications 4Solvent SelectionTubing Kit (G1160-68706) for solvent selection (4 solvents) with G1160A, degasser an

Pagina 48 - Column Selection

46 Agilent 1100 Series Valves4 Valve ApplicationsSolvent Selection

Pagina 49 - Tabl e 13 Capillaries

47Agilent 1100 Agilent 1100 Series ValvesUser’s GuideAgilent Technologies5PartsGeneral Parts 48Capillaries and Fittings for analytical and preparative

Pagina 50 - Solvent Selection

48 Agilent 1100 Series Valves5PartsGeneral PartsTabl e 17 General partsDescription part number• CAN cable, 1m long 5181-1519• DC-Can cable 5181-1533•

Pagina 51 - Table 15 Tubing

Agilent 1100 Series Valves 49Parts 5Tabl e 19 Rebuild kits and repair parts for Agilent 1100 Series ValvesDescription part number• Rebuilt kits and ro

Pagina 52 - 4 Valve Applications

50 Agilent 1100 Series Valves5PartsCapillaries and Fittings for analytical and preparative flow ratesTabl e 20 Capillary and tubing kitsDescription pa

Pagina 53 - Agilent Technologies

Agilent 1100 Series Valves 51Parts 5Capillaries and Fittings for micro valvesFor the operation of the Agilent 1100 Series Micro Valves we recommend th

Pagina 54 - General Parts

52 Agilent 1100 Series Valves5PartsTabl e 23 Fitting TypesFittings and ferrules Fitting typeABCDTabl e 24 PEEK coated fused silica capillaries for mic

Pagina 55 - Agilent 1100 Series Valves 49

Agilent 1100 Series Valves 53Parts 550 220 0.432 B/B G1375-8730150 280 0.550 C/D G1375-8730950 400 0.785 E/D G1315-6870350 500 0.982 C/D G1375-8730450

Pagina 56

ii Agilent 1100 Series Valvescontents4 Valve ApplicationsAlternating Column Regeneration 34Sample enrichment and Sample stripping 38Sample enrichment

Pagina 57

54 Agilent 1100 Series Valves5Parts

Pagina 58

55Agilent 1100 Agilent 1100 Series ValvesUser’s GuideAgilent Technologies6SpecificationsG1157A Agilent 1100 Series 2 position / 10 port valve 56G1158A

Pagina 59 - *seeTa b l e 2 3 on page 52

56 Agilent 1100 Series Valves6 SpecificationsTabl e 25 G1157A Agilent 1100 Series 2 position / 10 port valveLiquid contacts: Stainless Steel and PEEKP

Pagina 60 - 54 Agilent 1100 Series Valves

Agilent 1100 Series Valves 57Specifications 6* The G1159A 1100 Series 6 positions selection valve can be used at flow rates up to 100 ml/min, but with

Pagina 61 - Specifications

58 Agilent 1100 Series Valves6 SpecificationsVolume in flow passages: Stator (20° ports)27.2 nl, (45° ports) 30.5 nl, rotor seal 25.0 nl/grooveMaximum

Pagina 62 - 6 Specifications

59Agilent 1100 Agilent 1100 Series ValvesUser’s GuideAgilent Technologies7Safety InformationGeneral Safety Information 60Lithium Batteries Information

Pagina 63

60 Agilent 1100 Series Valves7 Safety InformationGeneral Safety InformationGeneral Safety InformationThe following general safety precautions must be

Pagina 64 - 58 Agilent 1100 Series Valves

Agilent 1100 Series Valves 61Safety Information 7General Safety Informationattempt internal service or adjustment unless another person, capable of re

Pagina 65 - Safety Information

62 Agilent 1100 Series Valves7 Safety InformationGeneral Safety InformationSafety SymbolsTable 31 shows safety symbols used on the instrument and in t

Pagina 66 - General Safety Information

Agilent 1100 Series Valves 63Safety Information 7Lithium Batteries InformationLithium Batteries InformationWARNINGDanger of explosion if battery is in

Pagina 67

1Agilent 1100 Agilent 1100 Series ValvesUser’s GuideAgilent Technologies1Installation and ConfigurationSite Requirements 2Unpacking the 1100 Series Va

Pagina 68 - Safety Symbols

64 Agilent 1100 Series Valves7 Safety InformationRadio InterferenceRadio InterferenceNever use cables other than the ones supplied by Aligent Technolo

Pagina 69 - Lithium Batteries Information

Agilent 1100 Series Valves 65Safety Information 7Sound EmissionSound EmissionManufacturer’s DeclarationThis statement is provided to comply with the r

Pagina 70 - Radio Interference

66 Agilent 1100 Series Valves7 Safety InformationSolvent InformationSolvent InformationObserve the following recommendations on the use of solvents.So

Pagina 71 - Sound Emission

Agilent 1100 Series Valves 67Safety Information 7Agilent Technologies on InternetAgilent Technologies on InternetFor the latest information on product

Pagina 72 - Solvent Information

68 Agilent 1100 Series Valves7 Safety InformationAgilent Technologies on Internet

Pagina 73

Agilent 1100 Series Valves 69IndexAaccessory kit, 3Agilent ChemStation, 8, 12Agilent on internet, 67alternating column regeneration, 34analytical scal

Pagina 74 - 7 Safety Information

70 Agilent 1100 Series ValvesIndexstator ring, 22, 24Status Indicators, 20switch failure, 30switching the valve, 16synchronization, 16, 32system error

Pagina 76 - 70 Agilent 1100 Series Valves

 Agilent Technologies 2002, 2004Printed in Germany 09/2004*G1156-90001**G1156-90001*G1156-90001www.agilent.comAgilent TechnologiesIn This BookThis Us

Pagina 77

2 Agilent 1100 Series Valves1 Installation and ConfigurationSite RequirementsSite RequirementsThe requirements for the Agilent 1100 Series Valves are

Pagina 78 - *G1156-90001*

Agilent 1100 Series Valves 3Installation and Configuration 1Unpacking the 1100 Series ValveUnpacking the 1100 Series ValveDamaged PackagingIf the deli