Agilent 1100 Series Valves 7
Installation and Configuration 1
Hardware Installation
Setting up the power connection to other Agilent 1100 Series Modules
The power for the Agilent 1100 Series External Valves can be supplied by one
of the following 1100 Series modules:
Agilent 1100 Series Preparative Pump (G1361A)
Agilent 1100 Series Fraction Collector (G1364A, G1364B, G1364C,
G1364D)
Agilent 1100 Series Well-Plate Autosampler (G1367A)
Agilent 1100 Series Micro Well-Plate Autosampler (G1377A)
Agilent 1100 Series Dual Loop Autosampler (G2258A)
Connect the DC-CAN Cable (PN 5181-1533) to the DC-CAN connector at the
back panel of the valve and to the DC-CAN connector on one of the Agilent
1100 Series modules listed above.
Setting up the power connection to the external power supply
If none of the Agilent 1100 Series modules from the list above is part of your
HPLC system, you have to supply the power through an external power supply
(P/N 0950-4422). Plug the DC-CAN adapter from the power supply into the
DC-CAN connector on the back panel of your 1100 Series External Valve.