32 Agilent 1100 Series Valves
3 Maintenance, Repair, and Troubleshooting
System Errors and Troubleshooting
Valve Synchronization
If the valve failed to switch or if the current position of the valve is unknown,
it might be necessary to synchronize the valve.
To synchronize the valve
Step
Note
1 Switch to the Diagnostics view.
2 Select Valve from the Maintenance menu. The valve should now display the Current
Position as displayed in Figure 18. If the
current position is 0, it is unknown
3 To synchronize the valve, choose any
position from the Synchronize to position
dropdown list and click OK.
Figure 18 Valve Synchronization