What do I need for proper cleaning? Cleaning Information
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
85
What do I need for proper
cleaning?
Some Standard Cleaning Equipment is necessary for cleaning your
instrument. For certain cleaning procedures, you may also require certain
Additional Cleaning Equipment.
Standard Cleaning Equipment
Before you can start your cleaning procedure you need the following
standard equipment:
Dust and shutter caps
Isopropyl alcohol
Cotton swabs
Soft tissues
Pipe cleaner
Compressed air
Dust and shutter caps
All of Agilent Technologies’ lightwave instruments are delivered with
either laser shutter caps or dust caps on the lightwave adapter. Any cables
come with covers to protect the cable ends from damage or
contamination.
We suggest these protective coverings should be kept on the equipment at
all times, except when your optical device is in use. Be careful when
replacing dust caps after use. Do not press the bottom of the cap onto the
fiber too hard, as any dust in the cap can scratch or pollute your fiber
surface.
If you need further dust caps, please contact your nearest Agilent
Technologies sales office.