Agilent Technologies
8922 Multi-Band Test System
Supplementary
User’s Guide
Systems Covered
HP/Agilent 8922P Multi-Band Test System
HP/Agilent 8922R Multi-Band Test System
HP/Agilent 8922X Multi-Band Test System
HP/Agilent 8922Y Multi-Band Test System
Agilent Part No. 08922-90226
Printed in UK
October 1998
Vista de pagina 0
1 2 3 4 5 6 ... 99 100

Indice de contenidos

Pagina 1 - User’s Guide

Agilent Technologies 8922 Multi-Band Test SystemSupplementaryUser’s GuideSystems CoveredHP/Agilent 8922P Multi-Band Test SystemHP/Agilent 8922R Multi-

Pagina 2

viiiAbout this GuideAbout this GuideOverviewThis User’s Guide is a supplement to the Manuals currently supplied with the HP/Agilent 8922 GSM Test Sets

Pagina 3 - Limitation of Warranty

Index86Procedureabsolute GSM BCH power level, 82RRAM Initialize, 8offset calibration value, 8Rear Panel Connections, 6return for repair, 77RF analyzer

Pagina 4 - Assistance

ContentsContents-1About this Guide viii1 Configuring the HP/Agilent 8922 Multi-Band Test System 1Before You Start 2Connection Overview 3Connection D

Pagina 5 - Statement of Compliance

ContentsContents-24 GPIB Commands 33Additional Commands to existing subsystems 34Dual Band Control Subsystem 36GPIB Programming Example Overview 39

Pagina 6 - Herstellerbescheinigung

ContentsContents-30.3 GMSK Data Recovery Specifications (Agilent 8922M only) 63FM Demodulation Output Specifications (Agilent 8922M only) 64Pulse Demodu

Pagina 8 - Safety Information

11Configuring the HP/Agilent 8922 Multi-BandTest SystemThis section provides information on connecting and setting up of the HP/Agilent 8922 Multi-Ban

Pagina 9 - Safety Symbols

2Configuring the HP/Agilent 8922 Multi-Band Test SystemBefore You StartBefore You StartBefore connecting the HP/Agilent 8922 or HP/Agilent 83220 to eac

Pagina 10 - About this Guide

3Configuring the HP/Agilent 8922 Multi-Band Test SystemConnection OverviewConnection OverviewConnecting the HP/Agilent 8922 to the HP/Agilent 83220 is

Pagina 11 - Contents

4Configuring the HP/Agilent 8922 Multi-Band Test SystemConnection DetailsConnection Details1. HP 8922 IF Link PortThe main port connecting the HP/Agile

Pagina 12

5Configuring the HP/Agilent 8922 Multi-Band Test SystemConnection Details4. GSM/DCS/PCS RF In/Out PortThis is the main device-under-test (DUT) connecti

Pagina 13

© Copyright 1998, Agilent Technologies. All rights reserved. Reproduction, adaptation, or translation without priorwritten permission is prohibited, e

Pagina 14

6Configuring the HP/Agilent 8922 Multi-Band Test SystemConnection Details8. RF In/Out Path DiagramThe diagram on the front panel of the HP/Agilent 8322

Pagina 15 - Test System

7Configuring the HP/Agilent 8922 Multi-Band Test SystemCalibration RoutineCalibration RoutineOnce the HP/Agilent 8922 Multi-Band Test System has been c

Pagina 16 - Before You Start

8Configuring the HP/Agilent 8922 Multi-Band Test SystemCalibration Routine12 Using the parts listed, follow the instructions given on the display.13 On

Pagina 17 - Connection Overview

9Configuring the HP/Agilent 8922 Multi-Band Test SystemZero the Power MeterZero the Power MeterThe HP/Agilent 8922 and HP/Agilent 83220 each have indep

Pagina 18 - Connection Details

10Configuring the HP/Agilent 8922 Multi-Band Test SystemZero the Power Meter4 Once selected, the power meter will automatically zero.To Zero the Power

Pagina 19 - 5. Coupled RF In/Out

11Configuring the HP/Agilent 8922 Multi-Band Test SystemZero the Power MeterChecking the Current Status of the Power MetersUse the following simple, ma

Pagina 20 - Rear Panel Connections

12Configuring the HP/Agilent 8922 Multi-Band Test SystemZero the Power Meter

Pagina 21 - Calibration Routine

132How to SectionThis section describes some of the basic procedures that may be required whileoperating the HP/Agilent 8922 Multi-Band Test System. F

Pagina 22 - Default Start Up Mode

14How to SectionPower-up and Operating Mode SelectionPower-up and Operating Mode SelectionThis section describes the procedure for powering up the HP/

Pagina 23 - Zero the Power Meter

15How to SectionEstablishing a CallEstablishing a CallThis procedure is identical to that of a Standard HP/Agilent 8922 Test System.1 Once the mobile

Pagina 24

iWarrantyWarrantyThis Agilent Technologies instrument product is warranted against defects in material andworkmanship for a period of one year from da

Pagina 25 - Set using a Dual-Band mobile:

16How to SectionMaking Measurements in the E-GSM900 BandMaking Measurements in the E-GSM900 BandThe procedures for making measurements in the E-GSM900

Pagina 26

17How to SectionChanging BandChanging BandAssignmentE-GSM900 to DCS18001 Select the Dual Band screen (Refer to Dual Band Control Screen on page 26 for

Pagina 27 - How to Section

18How to SectionChanging BandHandoverE-GSM900 to DCS18001 Select Dual Band Screen (Refer to Dual Band Control Screen on page 26 for moreinformation).2

Pagina 28 - Potential Problems

19How to SectionMaking Measurements in the DCS1800 BandMaking Measurements in the DCS1800 BandThe procedures for making measurements in the DCS1800 Ba

Pagina 29 - Establishing a Call

20How to SectionEnding A CallEnding A CallEnding a call is the same procedure as that used for a Standard HP/Agilent 8922Test System. Refer to the Agi

Pagina 30

213ScreensThe following section describes the screens that are part of the HP/Agilent 8922Multi-Band Test System. For more details on all other screen

Pagina 31 - Changing Band

22ScreensCell Configuration ScreenCell Configuration Screen1. Multi-band ReportThis field sets the number of neighbour cells, for each frequency band, w

Pagina 32 - Handover

23ScreensCell Status Screen - Active CellCell Status Screen - Active Cell1. Dual BandThis field is displayed whenever the system is operating in dual b

Pagina 33

24ScreensCell Status Screen - Test ModeCell Status Screen - Test Mode1. Dual Band, On/OffThis field appears when is selected, from the Cell Status scr

Pagina 34 - Ending A Call

25ScreensConfigure ScreenConfigure Screen1. CompatibleThis field selects the mode of operation of the HP/Agilent 8922. For moreinformation about this fie

Pagina 35 - 8922M⁄S User’s Guide

iiResponsibilities of the CustomerResponsibilities of the CustomerThe customer shall provide;1 Access to the products during the specified periods of

Pagina 36 - Cell Configuration Screen

26ScreensDual Band Control ScreenDual Band Control ScreenThis screen is used for performing inter-band channel changes while dual bandmobile testing.

Pagina 37 - 2a. TCH Ampl

27ScreensDual Band Control Screen2. Call StatusThis field displays the current status of the call. Refer to the Cell Status screendescribed in the Agil

Pagina 38 - 2. TCH Ampl OR Amplitude

28ScreensDual Band Control Screen5. Downlink AmplitudesThese 3 fields control the amplitudes of the downlink channels. These fieldsoperate differently d

Pagina 39 - Configure Screen

29ScreensDual Band Control Screen6. ExecuteThis field starts the process of moving the traffic channel from one band toanother. If no call is active, th

Pagina 40 - Dual Band Control Screen

30ScreensDual Band Control Screen9. TX LevelThis field tells the mobile which power level to use once a handover is requested.This field is linked to th

Pagina 41 - 4. Current Band

31ScreensError and Help ScreensError and Help ScreensError ScreenThis screen will be displayed if you try to access the Dual Band Control screenwithou

Pagina 42

32ScreensMS Information/Signaling ScreenMS Information/Signaling Screen1. Power ClassThis field displays the power class of the mobile. The display sho

Pagina 43 - 8. Level Control

334GPIB CommandsThe following section details the additional GPIB commands that are part of theHP/Agilent 8922 Multi-Band Test System. For description

Pagina 44 - 9. TX Level

34GPIB CommandsAdditional Commands to existing subsystemsAdditional Commands to existing subsystemsCell Configuration SubsystemDisplay SubsystemMS Inf

Pagina 45 - Error and Help Screens

35GPIB CommandsAdditional Commands to existing subsystemsPower Class DetailsThe commands for querying the power class can have different results depen

Pagina 46 - 2. MS Band

iiiNoticesNoticesThe material contained in this document is subject to change without notice. AGILENTTECHNOLOGIES MAKES NO WARRANTY OF ANY KIND WITH R

Pagina 47 - GPIB Commands

36GPIB CommandsDual Band Control SubsystemDual Band Control SubsystemDUALbandTCHEGSM?EGSMspintegerDCS?DCSspintegerTLEVelEGSM?EGSMspintegerDCS?DCSspint

Pagina 48 - MS Information Subsystem

37GPIB CommandsDual Band Control SubsystemE-GSM900 TCH CommandsThese commands set the parameters that will be used for an E-GSM900 trafficchannel after

Pagina 49 - Command Syntax Description

38GPIB CommandsDual Band Control SubsystemDual Band Control CommandsDUALband:AMPLitude:DTCH <real> Sets the DCS1800 TCH downlink amplitudeRefer

Pagina 50 - Dual Band Control Subsystem

39GPIB CommandsGPIB Programming Example OverviewGPIB Programming Example OverviewThe following example program performs a channel assignment, from the

Pagina 51 - DCS1800 TCH Commands

40GPIB CommandsExampleExample10 !---------------------------------------------------------------------!20 !

Pagina 52 - Dual Band Control Commands

41GPIB CommandsExample460 PAUSE470 !480 !490 !500 ! Setup the Mobile’s IMSI and then page the mobile510 !520 OUTPUT Addr;”MSINfo:PAGing:

Pagina 53 - Program Overview

42GPIB CommandsExample960 !970 !980 !990 ! Now send the assignment command1000 !1010 DISP “Execute multi-band assignment to DCS band”1020 O

Pagina 54

43GPIB CommandsExample1460 DISP “Execute multi-band assignment to GSM band”1470 OUTPUT Addr;”DUAL:TCHC:EXEC”1480 Try=01490 REPEAT1500 WAIT .215

Pagina 55

44GPIB CommandsGPIB Commands for Including Power Zeroing in an Automatic Test RoutineGPIB Commands for Including Power Zeroing in anAutomatic Test Rou

Pagina 56

455System SpecificationsThese specifications describe the system’s warranted performance and apply aftera 30 minute warm-up. These specifications are va

Pagina 57

ivElectromagnetic Compatibility (EMC) InformationElectromagnetic Compatibility (EMC) InformationThis product has been designed to meet the protection

Pagina 58 - Automatic Test Routine

46System SpecificationsSupplemental Characteristics (shown in italics)Supplemental Characteristics (shown in italics)These are intended to provide addi

Pagina 59 - System Specifications

47System SpecificationsGSM Functionality SpecificationsGSM Functionality SpecificationsBit/frame error rate measurements: Class Ia, Ib, and Class II bit

Pagina 60

48System SpecificationsRF Generator SpecificationsRF Generator SpecificationsRF In/Output connector FrequencyRange: 880 to 960 MHz and 1805 to 1990 MHzR

Pagina 61 - 1. Dual band operation only

49System SpecificationsRF Generator SpecificationsSupplemental characteristics in E-GSM900 mode when BCH is in GSM900band and TCH is in DCS1800 bandBCH

Pagina 62 - RF Generator Specifications

50System Specifications10 MHz to 1 GHz Aux RF Out Connector Specifications10 MHz to 1 GHz Aux RF Out Connector SpecificationsFrequencyRange: 10 MHz to 1

Pagina 63

51System Specifications1710 to 1990 MHz Aux RF Out Connector Specifications (HP/Agilent 83220A only)1710 to 1990 MHz Aux RF Out Connector Specifications

Pagina 64 - Frequency

52System Specifications0.3 GMSK Modulation Specifications0.3 GMSK Modulation SpecificationsAfter one timeslot, 577 µs, from an isolated RF generator tri

Pagina 65 - (HP/Agilent 83220A only)

53System SpecificationsPulse Modulation SpecificationsPulse Modulation SpecificationsInput levels (Agilent 8922M Only): TTLRise/fall time (10% to 90%):

Pagina 66 - Supplemental characteristics

54System Specifications30 dB Pulse Modulation Specifications (Agilent 8922M Only)30 dB Pulse Modulation Specifications(Agilent 8922M Only)All timeslots

Pagina 67

55System SpecificationsRF Analyzer SpecificationsRF Analyzer SpecificationsRF in/out connectorFrequency Range: 880 to 960 MHz and 1710 to 1990 MHzFreque

Pagina 68 - (Agilent 8922M Only)

vDeclaration of ConformityDeclaration of Conformity according to ISO/IEC Guide 22 and EN45014Manufacturer’s Name: Agilent TechnologiesSouth Queensferr

Pagina 69 - RF Analyzer Specifications

56System SpecificationsCW RF Frequency Measurement SpecificationsCW RF Frequency Measurement SpecificationsInput frequency setting error: ± 500 kHzAccur

Pagina 70 - System Specifications

57System SpecificationsCW RF Power Measurement Specifications (RF in/out) onlyCW RF Power Measurement Specifications(RF in/out) onlyInput frequency sett

Pagina 71 - (RF in/out) only

58System SpecificationsPeak Transmitter Carrier Power Measurement Specifications (RF in/out) onlyPeak Transmitter Carrier Power MeasurementSpecification

Pagina 72 - Supplemental Characteristics

59System SpecificationsPeak Transmitter Carrier Power Measurement Specifications (RF in/out) only1600 1700 1800 1900 2000 2100Frequency (MHz)0.60.40-0.2

Pagina 73

60System SpecificationsPulse on/off Ratio Measurement Specifications (requires option 006)Pulse on/off Ratio Measurement Specifications(requires option

Pagina 74 - (requires option 006)

61System SpecificationsAmplitude Envelope Measurement SpecificationsAmplitude Envelope Measurement SpecificationsAfter one timeslot, 577 µs from an isol

Pagina 75

62System SpecificationsPhase and Frequency Measurement SpecificationsPhase and Frequency Measurement SpecificationsAfter one timeslot, 577 µs from an is

Pagina 76

63System Specifications0.3 GMSK Data Recovery Specifications (Agilent 8922M only)0.3 GMSK Data Recovery Specifications(Agilent 8922M only)1After one tim

Pagina 77 - (Agilent 8922M only)

64System SpecificationsFM Demodulation Output Specifications (Agilent 8922M only)FM Demodulation Output Specifications(Agilent 8922M only)1Sensitivity:

Pagina 78

65System SpecificationsPulse Demodulation Output Specifications (Agilent 8922M only)Pulse Demodulation Output Specifications(Agilent 8922M only)1Input f

Pagina 79

viSafety InformationSafety InformationThe following safety precautions must be observed during all phases of operation, service,and repair of this ins

Pagina 80

66System SpecificationsOutput RF Spectrum Measurement Specifications (requires option 006)Output RF Spectrum Measurement Specifications(requires option

Pagina 81

67System SpecificationsSpectrum Analyzer Specifications (requires option 006)Spectrum Analyzer Specifications(requires option 006)Frequency span/resolut

Pagina 82 - Audio Source Specifications

68System SpecificationsAudio Source SpecificationsAudio Source SpecificationsFrequencyRange: DC to 25 kHzAccuracy: 0.025% of settingSupplemental charact

Pagina 83 - Audio Analyzer Specifications

69System SpecificationsAudio Analyzer SpecificationsAudio Analyzer SpecificationsFrequency measurementRange: 20 Hz to 400 kHzAccuracy: ± (0.02% + 1 coun

Pagina 84 - Audio detectors

70System SpecificationsAudio Analyzer SpecificationsDistortion measurementFundamental frequency: 1 kHz ± 5HzInput level range: 30 mVrms to 30 VrmsDispla

Pagina 85 - Oscilloscope Specifications

71System SpecificationsOscilloscope SpecificationsOscilloscope SpecificationsFrequency range (3 dB): 2 Hz to 50 kHzScale/division: 10 mV to 10 V in 1, 2

Pagina 86 - Remote programming

72System SpecificationsRemote programmingRemote programmingGPIB: IEEE Standard 488.2Functions implemented: SH1, AH1, T6, L4, SR1, RL1, LE0, TE0,PP0, D

Pagina 87 - Printer Support

73System SpecificationsPrinter SupportPrinter SupportRS-232: 3 wire RJ-11 connector used for serialdata transferParallel PortThis port is used with pri

Pagina 88 - General Specifications

74System SpecificationsGeneral SpecificationsGeneral SpecificationsSize: HP 8922 + HP 83220: 310H x 426W x574D mm (12.25 x 16.75 x 23 inch)Weight: HP/Ag

Pagina 89 - Reference Specifications

75System SpecificationsReference SpecificationsReference SpecificationsThe accuracy needs for testing GSM radios require the unit to be operated withthe

Pagina 90 - Reference Specifications

viiSafety SymbolsSafety SymbolsThe following symbols on the instrument and in the manual indicate precautions whichmust be taken to maintain safe oper

Pagina 91

76System SpecificationsReference Specifications

Pagina 92 - Return Procedure

776How To Return Equipment for Repair

Pagina 93 - Sales and Service Offices

78How To Return Equipment for RepairReturn ProcedureReturn ProcedureIf you need to return your HP/Agilent 8922 or HP/Agilent 83220 to AgilentTechnolog

Pagina 94 - United States:

79How To Return Equipment for RepairSales and Service OfficesSales and Service OfficesAny adjustment, maintenance, or repair of this product mustbe per

Pagina 95 - Power Level

80How To Return Equipment for RepairSales and Service OfficesCanadaAgilent Technologies Canada Inc.5150 Spectrum Way,Mississauga, OntarioL4W 5G1(tel) 1

Pagina 96 - Procedure

817Appendix A - Establishing Absolute GSM BCHPower Level

Pagina 97

82Appendix A - Establishing Absolute GSM BCH Power LevelProcedureProcedureThe following procedure can be used for determining the absolute GSM BCHpeak

Pagina 98 - To enhance this measurement

83Appendix A - Establishing Absolute GSM BCH Power LevelProcedure7 Set the Traffic Chan field to . This will remove the pulsing from theBCH channel.8

Pagina 99

84Appendix A - Establishing Absolute GSM BCH Power LevelProcedureTo enhance this measurement1 Press .2 Press softkey SWP TIME and change to MAN.3 Ente

Pagina 100

Index85Numerics1710-1990 MHz Aux RF In Port, 5AAbsolute GSM BCH power level, 82amplitude downlink, 28Assign/Handover, 26assignment, 17Aux RF In/Out Po