Agilent Technologies 81680A Guía de usuario Pagina 46

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 60
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 45
3-16
Reference
Cleaning Connections for Accurate Measurements
3
Dry the inside of the adapter with a clean, dry, foam swab.
4
Blow through the adapter using filtered, dry, compressed air.
Nitrogen gas or compressed dust remover can also be used. Do not shake, tip,
or invert compressed air canisters, because this releases particles in the can
into the air. Refer to instructions provided on the compressed air canister.
Vista de pagina 45
1 2 ... 41 42 43 44 45 46 47 48 49 50 51 ... 59 60

Comentarios a estos manuales

Sin comentarios