Chapter 3 165
Screen Reference
CONFIGURATION
3. Screen Reference
How to Update the Firmware
If Update is set to On and a firmware update card is inserted in the MEMORY
CARD slot, the version number of the firmware on the card is displayed in the
New Firmware field. If a firmware update card is not inserted or an improper card
is inserted, “?.??.??” is shown in this field.
Figure 3-65 Firmware Update Screen (1/2)
Softkey Menu
o Yes: Initiates the updating process. When this softkey is pressed with the
proper update card inserted, the screen shown in Figure 3-66 is displayed.
o No: Changes from Update: On to Update: Off without updating the current
firmware.
Figure 3-66 Firmware Update Screen (2/2)
Softkey Menu
o OK: Begins updating the firmware to the new revision. Following the
update card #1, it is required to insert the update card #2 according to the
operation message.
After completing the update process, the Test Set reboot to display the
initial screen.