5-60 N9360A Multi UE Tester W-CDMA User Manual
5 Screen Reference
RF Test in Manual Test
Table 5-17 describes the measurement items of RF test in
Manual Test. The limits for measurement items can be set on the
Configuration: Test Condition screen. Refer to Table 5-46,
“[Configuration: Test Sequence] Screen Input Field,” on
page 5-153.
TEST USIM
Insert the TEST USIM provided by Agilent in the mobile phone
before performing any test. The tester will not be able to
perform measurements with a USIM provided by cell phone
operator or any other TEST USIMs.
Table 5-17 Description of Manual Test Measurement Items
Measurement Item Description
TX Power Measures the TX power.
Frequency Error Measures the frequency error.
EVM Measures the error vector magnitude.
Origin Offset Measures the origin offset.
BER Measures the bit error rate with loopback using
downlink and uplink.
The default state at power on is the last setting stored in the internal
memory of the Tester before power off.
For numeric value entry, the changing magnification softkey menu is
available. Refer to “Storing Numeric Values" on page 4-10 and “Changing
Magnification Softkey" on page 4-13.