Agilent TechnologiesAgilent Technologies5500 Scanning Probe MicroscopeUser’s Guide

Pagina 2

Agilent 5500 SPM User’s Guide 10ContentsTable of Co ntentsRead This First Specifications 4Laser Safety Information 4Power Supply 5

Pagina 3

Contact Mode Imaging 5Agilent 5500 SPM User’s Guide 100two components. Comparing the friction and topography images helps to differentiate the impact

Pagina 4 - Laser Safety Information

Contact Mode Imaging 5Agilent 5500 SPM User’s Guide 101tip. Higher force can place undue wear on the tip and, in the extreme, can damage the tip or sa

Pagina 5 - Power Supply

Contact Mode Imaging 5Agilent 5500 SPM User’s Guide 102speed will be 2-5 lines/second for smooth surfaces. For rougher surfaces a lower scan speed may

Pagina 6 - Disclaimers

103Agilent 5500 SPMUser’s GuideAgilent Technologies6AC ModesAcoustic AC Mode (AAC) 104AAC Mode 104Constant Height Mode 109Magnetic AC (MAC) M

Pagina 7

AC Modes 6Agilent 5500 SPM User’s Guide 104Acoustic AC Mode (AAC)In Acoustic AC (AAC) Mode AFM, a piezo-electric transducer in the nose assembly drive

Pagina 8 - Declaration of Conformity

AC Modes 6Agilent 5500 SPM User’s Guide 1055Choose Controls > AC Mode to open the ACAFM Controls window (Figure 72).Figure 72 ACAFM Controls window

Pagina 9 - Contact Information

AC Modes 6Agilent 5500 SPM User’s Guide 106Figure 73 AC Mode Tune window Figure 74 AC Tune window with resonance peak at ~154.4 kHz10 The next step is

Pagina 10 - 2 Agilent 5500 SPM Components

AC Modes 6Agilent 5500 SPM User’s Guide 107storage box should indicate the range in which the primary resonance frequency will be found.11 In the uppe

Pagina 11 - 4 Preparing for Imaging

AC Modes 6Agilent 5500 SPM User’s Guide 108Next, bring the tip into contact with the sample. In AAC mode, “contact” occurs when the cantilever oscilla

Pagina 12 - 7 Additional Imaging Modes

AC Modes 6Agilent 5500 SPM User’s Guide 10925In the Realtime Images window, make sure that Topography and Deflection are displayed.26 In the Scan and

Pagina 13 - 9 Closed-Loop Scanners

Agilent 5500 SPM User’s Guide 11ContentsTwo-Piece Nose Assemblies 36Scanner 38Detector 40Sample Plates 41Video System 43Head

Pagina 14 - 13 Temperature Control

AC Modes 6Agilent 5500 SPM User’s Guide 110response to amplitude changes. This lack of feedback reduces signal noise, enabling high resolution imaging

Pagina 15 - Index

AC Modes 6Agilent 5500 SPM User’s Guide 111MAC III offers additional lock-in amplifiers for other more complex modes as well). Specially coated MAC ca

Pagina 16

AC Modes 6Agilent 5500 SPM User’s Guide 1122In the ACAFM Controls dialog box, choose MAC as the Drive Mechanism. Top MAC ModeIn Top MAC Mode AFM, the

Pagina 17

AC Modes 6Agilent 5500 SPM User’s Guide 113well-defined resonant peak in MAC Mode makes the method particularly effective.Q Control uses a feedback lo

Pagina 18 - SPM Basics

114Agilent 5500 SPMUser’s GuideAgilent Technologies7Additional Imaging ModesScanning Tunneling Microscopy (STM) 114Current Sensing AFM (CSAFM) 1

Pagina 19

Additional Imaging Modes 7Agilent 5500 SPM User’s Guide 115available with three different preamplifiers for varying sensitivity (Table 2). Figure 80 S

Pagina 20 - SPM Techniques

Additional Imaging Modes 7Agilent 5500 SPM User’s Guide 116located beneath the tip, can be field-replaced to adjust the sensitivity if necessary.Figur

Pagina 21 - Atomic Force Microscopy (AFM)

Additional Imaging Modes 7Agilent 5500 SPM User’s Guide 1175Attach an electrode from the sample plate to the sample. Lift the spring-loaded electrode

Pagina 22

Additional Imaging Modes 7Agilent 5500 SPM User’s Guide 118Figure 84 Servo window settings for STM imaging10 Enter the Setpoint current, in nanoamps,

Pagina 23 - Contact Mode AFM

Additional Imaging Modes 7Agilent 5500 SPM User’s Guide 119Current Sensing AFM (CSAFM)In Current Sensing AFM (CSAFM) an ultra-sharp AFM cantilever, co

Pagina 24 - Intermittent Contact AFM

Agilent 5500 SPM User’s Guide 12ContentsRemoving the Body of the Two-Piece Nose Assembly 68Inserting a Probe in the Two-Piece Nose Assembly 69Insertin

Pagina 25 - Acoustic AC (AAC) AFM

Additional Imaging Modes 7Agilent 5500 SPM User’s Guide 1201 nA/V (blue) or 0.1 nA/V (green). See Table 2 on page 115 for more details.Figure 86 CSAFM

Pagina 26

Additional Imaging Modes 7Agilent 5500 SPM User’s Guide 1216In PicoView, choose Mode > CSAFM.7 In the Servo window enter the Bias Voltage. Typical

Pagina 27 - Current Sensing Mode (CSAFM)

Additional Imaging Modes 7Agilent 5500 SPM User’s Guide 12214Watching the video system, bring the tip and sample very close to contact:a Adjust the fo

Pagina 28

Additional Imaging Modes 7Agilent 5500 SPM User’s Guide 123Lateral Force Microscopy (LFM)Lateral Force Microscopy is a derivative of Contact Mode. Dur

Pagina 29

Additional Imaging Modes 7Agilent 5500 SPM User’s Guide 124NOTEIt is important in LFM that the LFM signal on the Head Electronics Box be carefully set

Pagina 30

Additional Imaging Modes 7Agilent 5500 SPM User’s Guide 125Dynamic Lateral Force Microscopy (DLFM)In Dynamic Lateral Force Microscopy (DLFM), a lead z

Pagina 31

Additional Imaging Modes 7Agilent 5500 SPM User’s Guide 1266Bring the tip close to the sample:a Press the Close switch on the HEB to raise the sample

Pagina 32 - Agilent 5500 SPM Components

Additional Imaging Modes 7Agilent 5500 SPM User’s Guide 12712In the Scan and Motor window’s Scan tab, enter:a Scan Speed of 1-2 ln/s.b Resolution of 2

Pagina 33 - Figure 10

Additional Imaging Modes 7Agilent 5500 SPM User’s Guide 128For the MAC III controller these connections are made in software.To image in Force Modulat

Pagina 34

Additional Imaging Modes 7Agilent 5500 SPM User’s Guide 1299Now set up the additional AC oscillation:a Choose Controls > AC Mode to open the ACAFM

Pagina 35 - Agilent 5500 SPM Components 2

Agilent 5500 SPM User’s Guide 13ContentsElectrostatic Force Microscopy (EFM) 130Kelvin Force Microscopy (KFM) 1348 Scanner Maintenance and C

Pagina 36

Additional Imaging Modes 7Agilent 5500 SPM User’s Guide 130Electrostatic Force Microscopy (EFM)Electrostatic Force Microscopy (EFM) is a qualitative m

Pagina 37 - Mode/Contact Mode imaging

Additional Imaging Modes 7Agilent 5500 SPM User’s Guide 1316Choose Controls > Advanced > AC Mode. The EFM Controls window will open (Figure 88).

Pagina 38

Additional Imaging Modes 7Agilent 5500 SPM User’s Guide 13210Bring the tip close to contact with the sample:a Press the Close switch on the HEB to rai

Pagina 39

Additional Imaging Modes 7Agilent 5500 SPM User’s Guide 133amplitude exceeding 10 V, beyond which the signal will be clipped. The default value is x1

Pagina 40 - Detector

Additional Imaging Modes 7Agilent 5500 SPM User’s Guide 134Kelvin Force Microscopy (KFM)Kelvin Force Microscopy (KFM) is similar to EFM. An additional

Pagina 41 - Sample Plates

Additional Imaging Modes 7Agilent 5500 SPM User’s Guide 1354Choose Controls > Spectroscopy to open the Spectroscopy window (Figure 90). Figure 90 U

Pagina 42

Additional Imaging Modes 7Agilent 5500 SPM User’s Guide 136For more on advanced options for KFM Mode see Chapter 11.

Pagina 43 - Video System

137Agilent 5500 SPMUser’s GuideAgilent Technologies8Scanner Maintenance and CalibrationCare and Handling of the Probes and Scanner 138Probes 138

Pagina 44 - Head Electronics Box (HEB)

Scanner Maintenance and Calibration 8Agilent 5500 SPM User’s Guide 138Care and Handling of the Probes and ScannerProbesAlways store probes at room tem

Pagina 45 - AFM Controller

Scanner Maintenance and Calibration 8Agilent 5500 SPM User’s Guide 139nose assembly will likely need to be replaced. Use careful handling to avoid dam

Pagina 46 - Vibration Isolation Chamber

Agilent 5500 SPM User’s Guide 14Contents10 MAC ModeList of MAC Mode Components 162Connections 163Hardware and Sample Setup 16411 MAC II

Pagina 47 - Software

Scanner Maintenance and Calibration 8Agilent 5500 SPM User’s Guide 140explained below. Though they are explained separately for simplicity, they may n

Pagina 48 - System Options

Scanner Maintenance and Calibration 8Agilent 5500 SPM User’s Guide 141by an equal and opposite field in the other direction. The effect of hysteresis

Pagina 49

Scanner Maintenance and Calibration 8Agilent 5500 SPM User’s Guide 142movement) and a single tube for Z motion. Since the range is small with these sc

Pagina 50 - Environmental Chamber

Scanner Maintenance and Calibration 8Agilent 5500 SPM User’s Guide 143Calibrating the Multi-Purpose ScannerRegular calibration ensures that Agilent mu

Pagina 51 - PicoTREC

Scanner Maintenance and Calibration 8Agilent 5500 SPM User’s Guide 144X CalibrationIn the Realtime Images window set up two Topography images, one for

Pagina 52 - PicoLITH

Scanner Maintenance and Calibration 8Agilent 5500 SPM User’s Guide 145X Non-LinearityTo check X non-linearity, in the Realtime Images window choose To

Pagina 53

Scanner Maintenance and Calibration 8Agilent 5500 SPM User’s Guide 146After updating, re-scan the calibration target. The spacing between features sho

Pagina 54

Scanner Maintenance and Calibration 8Agilent 5500 SPM User’s Guide 147X SensitivityUsing the Cross-section tool, measure the length of a set of featur

Pagina 55 - Facility Requirements

Scanner Maintenance and Calibration 8Agilent 5500 SPM User’s Guide 148the slow scanning axis so the range will be reduced as a time consideration. O

Pagina 56 - Utilities

Scanner Maintenance and Calibration 8Agilent 5500 SPM User’s Guide 149If the dimensions are not identical for the two features, adjust the Y Non-linea

Pagina 57

Agilent 5500 SPM User’s Guide 15ContentsImaging 207Tips for Temperature Controlled Imaging 20814 Environmental ControlEnvironmental Chamber

Pagina 58 - Connecting the Components

Scanner Maintenance and Calibration 8Agilent 5500 SPM User’s Guide 150on the edge of a feature while the scan was moving upwards. The red marker was p

Pagina 59 - Preparing for Imaging

Scanner Maintenance and Calibration 8Agilent 5500 SPM User’s Guide 151If the measured dimension does not match the actual, the adjust the Y Sensitivit

Pagina 60 - One-Piece Nose Assembly

Scanner Maintenance and Calibration 8Agilent 5500 SPM User’s Guide 152top and the bottom of the data plot is flat before making the measurement.If the

Pagina 61 - Preparing for Imaging 4

153Agilent 5500 SPMUser’s GuideAgilent Technologies9Closed-Loop ScannersScanner Types 153Z-Axis Closed-Loop Scanner 153X/Y/Z Closed-Loop Scanner

Pagina 62

Closed-Loop Scanners 9Agilent 5500 SPM User’s Guide 154Positional drift that may be present in an open-loop system is continuously corrected with the

Pagina 63

Closed-Loop Scanners 9Agilent 5500 SPM User’s Guide 155Figure 106 Scanner Setup window4 The values shown in Figure 106 for the X, Y and Z Sensor Offse

Pagina 64

Closed-Loop Scanners 9Agilent 5500 SPM User’s Guide 1567In the Scan and Motor window enter:a A very large number (e.g., 9999) for the Scan Size. It wi

Pagina 65

Closed-Loop Scanners 9Agilent 5500 SPM User’s Guide 15711Adjust the X Sensor values:a If the line slopes down from the upper left part of the graph, s

Pagina 66

Closed-Loop Scanners 9Agilent 5500 SPM User’s Guide 15817Adjust the Sensitivity of the X sensor:a In the Realtime Images window choose Tools > Hori

Pagina 67 - Two-Piece Nose Assembly

Closed-Loop Scanners 9Agilent 5500 SPM User’s Guide 159of the Z piezo is being measured, adjusting the Z Sensor Gain and Offset should be done without

Pagina 68

16Agilent 5500 SPMUser’s GuideAgilent Technologies1Introduction to the Agilent 5500Overview of Agilent SPM System 17SPM Basics 18SPM Techniques

Pagina 69

Closed-Loop Scanners 9Agilent 5500 SPM User’s Guide 160The goal of the calibration procedure is to make the Z Sensor output appear as in Figure 110: F

Pagina 70

Closed-Loop Scanners 9Agilent 5500 SPM User’s Guide 161Once the graph looks like Figure 110, the Offset and Gain are properly set and the Z Sensor is

Pagina 71

162Agilent 5500 SPMUser’s GuideAgilent Technologies10MAC ModeList of MAC Mode Components 162Connections 163Hardware and Sample Setup 164The M

Pagina 72 - Aligning the Laser

MAC Mode 10Agilent 5500 SPM User’s Guide 163Please contact Agilent if any of these items are missing.ConnectionsThe MAC Mode controller is shown in Fi

Pagina 73

MAC Mode 10Agilent 5500 SPM User’s Guide 1643Input summed to AAC drive signal.4 Deflection output from detector.5 Amplitude output from lock-in amplif

Pagina 74

MAC Mode 10Agilent 5500 SPM User’s Guide 165Chapter 6, “AC Modes.” Please refer to Chapter 6 for more on how to set up the microscope for imaging.In A

Pagina 75

166Agilent 5500 SPMUser’s GuideAgilent Technologies11MAC III ModeInitial Setup 167List of MAC III Components 167Connections 168Hardware and S

Pagina 76

MAC III Mode 11Agilent 5500 SPM User’s Guide 167Not only is this arrangement very time efficient, it also ensures that the data is extremely reliable

Pagina 77 - Figure 52 Laser alignment

MAC III Mode 11Agilent 5500 SPM User’s Guide 168Please contact Agilent if any of these items are missing.ConnectionsThe MAC III controller is shown in

Pagina 78

MAC III Mode 11Agilent 5500 SPM User’s Guide 169In addition to the standard cabling for your microscope, the following connections must be made to use

Pagina 79

Introduction to the Agilent 5500 1Agilent 5500 SPM User’s Guide 17Overview of Agilent SPM System The main component of the Agilent 5500 SPM system is

Pagina 80

MAC III Mode 11Agilent 5500 SPM User’s Guide 170cable between the MICROSCOPE connector on the HEB and the 44-pin connector on the microscope base.AFM

Pagina 81

MAC III Mode 11Agilent 5500 SPM User’s Guide 171summed in to any or all of the drives by using Sum External Drive on the Lock-in tab of the Advanced A

Pagina 82

MAC III Mode 11Agilent 5500 SPM User’s Guide 172automatically adjust parameters to appropriate values. The changes will also be visible in the Advance

Pagina 83 - Mounting the Sample

MAC III Mode 11Agilent 5500 SPM User’s Guide 173The following parameters are available in the ACAFM Controls window:Amplitude Displays the amplitude,

Pagina 84

MAC III Mode 11Agilent 5500 SPM User’s Guide 174alignment procedure the Pass Through boxes will be selected to allow the correct signals to pass throu

Pagina 85

MAC III Mode 11Agilent 5500 SPM User’s Guide 175Amplitude Displays the amplitude, in volts, of cantilever lateral oscillation amplitude.Drive (%) The

Pagina 86 - Using the Video System

MAC III Mode 11Agilent 5500 SPM User’s Guide 176cantilever modulation that results from this applied signal is monitored as a measure of the sample’s

Pagina 87

MAC III Mode 11Agilent 5500 SPM User’s Guide 17710 V, beyond which the signal will be clipped. The default value is x1 (the amplitude times 1).Drive M

Pagina 88

MAC III Mode 11Agilent 5500 SPM User’s Guide 178Choose Mode > EFM to open the EFM Controls window:Figure 120 EFM Controls windowThe Main tab includ

Pagina 89

MAC III Mode 11Agilent 5500 SPM User’s Guide 17910 V, beyond which the signal will be clipped. The default value is x1 (the amplitude times 1).Zero Ph

Pagina 90

Introduction to the Agilent 5500 1Agilent 5500 SPM User’s Guide 18SPM BasicsScanning Probe Microscopy (SPM) is a large and growing collection of techn

Pagina 91

MAC III Mode 11Agilent 5500 SPM User’s Guide 180Optimize Phase shifts the phase signal to maximize the X Component 2 (i.e., to maximize contrast).X Co

Pagina 92 - Contact Mode Imaging

MAC III Mode 11Agilent 5500 SPM User’s Guide 181The Main tab shows settings for Lock-in 1 and Q-Control:Amplitude Displays the amplitude, in volts, of

Pagina 93 - Constant Force Mode

MAC III Mode 11Agilent 5500 SPM User’s Guide 182mechanical response provided by Lock-in 1 and to see if there are any other resonances present. To do

Pagina 94

MAC III Mode 11Agilent 5500 SPM User’s Guide 183AC Mode window includes several tabs, each of which is described below.Lock-In TabsEach of the three l

Pagina 95

MAC III Mode 11Agilent 5500 SPM User’s Guide 184an amplitude exceeding 10 V, beyond which the signal will be clipped. The default value is x1 (the amp

Pagina 96

MAC III Mode 11Agilent 5500 SPM User’s Guide 185respect to the drive signal (they will remain orthogonal to each other). This feature is useful, for e

Pagina 97

MAC III Mode 11Agilent 5500 SPM User’s Guide 186Drive Out Routes the output from one of the three lock-in signals to the circuit controlling oscillati

Pagina 98

MAC III Mode 11Agilent 5500 SPM User’s Guide 187the box to replace the Ref Set value with the lock-in signal. Select Sum plus GND to pass the Ref Set

Pagina 99

MAC III Mode 11Agilent 5500 SPM User’s Guide 188Other TabThe Other tab includes additional parameters that control MAC III operation: Figure 124 Advan

Pagina 100 - Fine-Tuning the Image

MAC III Mode 11Agilent 5500 SPM User’s Guide 189the Q Control feedback loop. By default, Q Control is turned Off.Drive (%) Sets the amplitude of the Q

Pagina 101 - Scan Settings

Introduction to the Agilent 5500 1Agilent 5500 SPM User’s Guide 19collection of data points is then synthesized into the “SPM image,” a 3-dimensional

Pagina 102 - Contact Mode Imaging 5

190Agilent 5500 SPMUser’s GuideAgilent Technologies12Liquid CellLiquid Cell with Standard Sample Plate 191Liquid Cell with MAC Mode 193Flow-Thro

Pagina 103 - AC Modes

Liquid Cell 12Agilent 5500 SPM User’s Guide 191container; therefore, the sample must be large enough for the O-ring to seat.Figure 125 Liquid cell com

Pagina 104 - Acoustic AC Mode (AAC)

Liquid Cell 12Agilent 5500 SPM User’s Guide 192sample are almost touching (i.e., both the tip and sample are close to focus).4 In PicoView, click the

Pagina 105

Liquid Cell 12Agilent 5500 SPM User’s Guide 19317Because of the large movements involved in placing and aligning the liquid cell, you will likely need

Pagina 106 - AC Modes 6

194Agilent 5500 SPMUser’s GuideAgilent Technologies13Temperature ControlCantilevers for Temperature Controlled Imaging 194High Temperature Sample P

Pagina 107

Temperature Control 13Agilent 5500 SPM User’s Guide 195due to the difference in thermal expansion of the coated and uncoated sides. The bending may ad

Pagina 108 - and/or Y Offset values to

Temperature Control 13Agilent 5500 SPM User’s Guide 196The Lakeshore 332 Temperature Controller (Figure 129) drives the high temperature plates.Figure

Pagina 109

Temperature Control 13Agilent 5500 SPM User’s Guide 197ConnectionsFigure 130 shows the three cables included with both high temperature sample plates:

Pagina 110 - Magnetic AC (MAC) Mode

Temperature Control 13Agilent 5500 SPM User’s Guide 198connector parallel to the cable and pass it through a port in the chamber. Then screw the round

Pagina 111 - Standard MAC Mode

Temperature Control 13Agilent 5500 SPM User’s Guide 199Figure 132 Hot sample plate wiring diagramFigure 133 Hot MAC sample plate wiring diagram

Pagina 112 - Q Control

Notices© Agilent Technologies, Inc. 2008No part of this manual may be reproduced in any form or by any means (including electronic storage and retriev

Pagina 113

Introduction to the Agilent 5500 1Agilent 5500 SPM User’s Guide 20SPM TechniquesScanning Tunneling Microscopy (STM) The earliest, widely-adopted SPM t

Pagina 114 - Additional Imaging Modes

Temperature Control 13Agilent 5500 SPM User’s Guide 200Note that Cable 3 includes a tab on the black-wire side of the connector (Figure 134). The tab

Pagina 115 -  1 G 100 M

Temperature Control 13Agilent 5500 SPM User’s Guide 2019Press the Heater Range button and select Low, Medium or High.10 Press the Setpoint button and

Pagina 116

Temperature Control 13Agilent 5500 SPM User’s Guide 202Peltier (Cold MAC) Sample PlateThe Peltier Cold MAC sample plate lets you image in contact, AAC

Pagina 117

Temperature Control 13Agilent 5500 SPM User’s Guide 203The Lakeshore 332 Temperature Controller (Figure 129) is also used with the Peltier plate. The

Pagina 118

Temperature Control 13Agilent 5500 SPM User’s Guide 204ConnectionsFigure 137 shows the three cables included with the Peltier sample plate:Figure 137

Pagina 119 - Current Sensing AFM (CSAFM)

Temperature Control 13Agilent 5500 SPM User’s Guide 205Figure 139 shows the required wiring for the Peltier sample plate. The connection at the end of

Pagina 120

Temperature Control 13Agilent 5500 SPM User’s Guide 206Note that Cable 3 includes a tab on the black-wire side of the connector (Figure 140). The tab

Pagina 121

Temperature Control 13Agilent 5500 SPM User’s Guide 207Figure 141 Gravity-fed water cooling system for Peltier sample plate.Imaging1 Set up the micros

Pagina 122

Temperature Control 13Agilent 5500 SPM User’s Guide 20811On the Lakeshore controller press Heater Range and select Low.12 Press the Setpoint button an

Pagina 123 - Additional Imaging Modes 7

209Agilent 5500 SPMUser’s GuideAgilent Technologies14Environmental ControlEnvironmental Chamber 209Glove Box 212In addition to the vibration iso

Pagina 124

Introduction to the Agilent 5500 1Agilent 5500 SPM User’s Guide 21topography. Because the feedback response requires time, constant current mode is ty

Pagina 125

Environmental Control 14Agilent 5500 SPM User’s Guide 210atmospheric pressure and is not intended to provide a vacuum or high pressure environment.Fig

Pagina 126

Environmental Control 14Agilent 5500 SPM User’s Guide 211microscope base will fit into grooves in the environmental chamber base.5 Swing the microscop

Pagina 127

Environmental Control 14Agilent 5500 SPM User’s Guide 212environmental chamber through a hole in the side of the vibration chamber.Glove BoxThe glove

Pagina 128

213Agilent 5500 SPMUser’s GuideAgilent Technologies15ElectrochemistryEquipment 216Liquid Cell 216Electrodes 216Working Electrode and Pogo Ele

Pagina 129

Electrochemistry 15Agilent 5500 SPM User’s Guide 214electrochemistry experiments can be conducted using either AFM Modes or STM Mode.NOTEAs of the wri

Pagina 130

Electrochemistry 15Agilent 5500 SPM User’s Guide 215Figure 145 Electrochemistry experiment schematicFigure 146 Electrochemistry experimental setup usi

Pagina 131

Electrochemistry 15Agilent 5500 SPM User’s Guide 216EquipmentThe equipment needed to perform electrochemistry experiments can be as simple as a liquid

Pagina 132

Electrochemistry 15Agilent 5500 SPM User’s Guide 217Reference ElectrodeThe reference electrode (RE) should have a diameter of 0.51 mm (0.02 in). It wi

Pagina 133

Electrochemistry 15Agilent 5500 SPM User’s Guide 218CleaningThorough cleaning of all components will greatly improve the results of electrochemistry e

Pagina 134 - Kelvin Force Microscopy (KFM)

Electrochemistry 15Agilent 5500 SPM User’s Guide 219Electrode CleaningElectrodes should be carefully cleaned prior to assembling the liquid cell. This

Pagina 135

Introduction to the Agilent 5500 1Agilent 5500 SPM User’s Guide 22detector. As the cantilever bends, the position of the laser spot changes. The resul

Pagina 136

Electrochemistry 15Agilent 5500 SPM User’s Guide 2208Insert the counter and reference electrodes into the sample plate block (Figure 146). Push the sp

Pagina 137 - Agilent Technologies

Electrochemistry 15Agilent 5500 SPM User’s Guide 221operational amplifier, but the tip itself is not connected physically to ground.2 In AFM Mode the

Pagina 138 - Nose Assembly

Electrochemistry 15Agilent 5500 SPM User’s Guide 222bias, probe potential and sample potential. Any one of these potentials can be swept during cyclic

Pagina 139 - Scanner Characteristics

223Agilent 5500 SPMUser’s GuideAgilent TechnologiesAppendix AWiring DiagramsAgilent 5500 SPM Standard Wiring Diagram 224Agilent 5500 SPM with MAC M

Pagina 140 - Hysteresis

224 Agilent 5500 SPM User’s GuideA Wiring DiagramsAgilent 5500 SPM Standard Wiring DiagramFigure 150 Wiring diagram for Agilent 5500 standard configur

Pagina 141 - Other Characteristics

225 Agilent 5500 SPM User’s GuideA Wiring DiagramsAgilent 5500 SPM with MAC Mode ControllerFigure 151 Wiring diagram for Agilent 5500 SPM with MAC Mod

Pagina 142 - Figure 93)

226 Agilent 5500 SPM User’s GuideA Wiring DiagramsAgilent 5500 SPM with MAC Mode, Force Modulation ImagingFigure 152 Wiring diagram for Agilent 5500 S

Pagina 143

227 Agilent 5500 SPM User’s GuideA Wiring DiagramsAgilent 5500 SPM with MAC III OptionFigure 153 Wiring diagram for Agilent 5500 SPM with MAC III Opti

Pagina 144 - X Calibration

228 Agilent 5500 SPM User’s GuideA Wiring DiagramsAgilent 5500 SPM with MAC III Option and Closed Loop ScannerFigure 154 Wiring diagram for Agilent 55

Pagina 145 - X Non-Linearity

IndexAgilent 5500 SPM User’s Guide 229IndexAAAC Mode, 103, 130AC Mode, 24, 103, 113Acoustic, 25, 103, 104, 130Constant height, 109Magnetic, 26, 1

Pagina 146 - X Hysteresis

Introduction to the Agilent 5500 1Agilent 5500 SPM User’s Guide 23described below, can be generally described by their function within these three dom

Pagina 147 - Y Calibration

IndexAgilent 5500 SPM User’s Guide 230GGains, 96, 108, 154Optimizing, 101Glove box, 212, 219HHead Electronics Box, 81, 93, 124, 162, 167, 221HEB,

Pagina 148 - Y Non-Linearity

IndexAgilent 5500 SPM User’s Guide 231Reference electrode, 217, 220RequirementsAcoustic noise, 56Resolution, 97Retrace, 141, 144SSample plate, 21

Pagina 149 - Y Hysteresis

Agilent Technologies5500 SPM User’s GuidePart Number N9410-90001© Agilent Technologies, Inc. 2008

Pagina 150 - Y Sensitivity

Introduction to the Agilent 5500 1Agilent 5500 SPM User’s Guide 24tip makes gentle contact with the sample, exerting from ~0.1-1000 nN force on the sa

Pagina 151 - Z Calibration

Introduction to the Agilent 5500 1Agilent 5500 SPM User’s Guide 25components.The complex tip-sample forces cause changes in the amplitude, phase and r

Pagina 152 - Archive the Calibration Files

Introduction to the Agilent 5500 1Agilent 5500 SPM User’s Guide 26amplitude—this reduction is used as a feedback signal to maintain constant amplitude

Pagina 153 - Closed-Loop Scanners

Introduction to the Agilent 5500 1Agilent 5500 SPM User’s Guide 27Figure 8 Magnetic AC mode (MAC mode)Top MAC ModeIn standard MAC Mode the magnetic co

Pagina 154 - Calibration

Introduction to the Agilent 5500 1Agilent 5500 SPM User’s Guide 28between the conducting cantilever and sample, a current is generated which is used t

Pagina 155

Introduction to the Agilent 5500 1Agilent 5500 SPM User’s Guide 29Lateral Force Microscopy (LFM)Lateral Force Microscopy (LFM) is a derivative of Cont

Pagina 156 - set to 0.000

Read This FirstN9410-90001 Agilent 5500 SPM User’s Guide 3Read This First Warranty Agilent warrants Agilent hardware, accessories and supplies against

Pagina 157

Introduction to the Agilent 5500 1Agilent 5500 SPM User’s Guide 30cantilever due to interatomic magnetic force that persists for greater tip-sample se

Pagina 158 - Z Sensor Calibration

Introduction to the Agilent 5500 1Agilent 5500 SPM User’s Guide 31provides a quantitative analysis of changes in the applied or intrinsic electrostati

Pagina 159

32Agilent 5500 SPMUser’s GuideAgilent Technologies2Agilent 5500 SPM ComponentsMicroscope 34Probes 35Nose Assembly 36One-Piece Nose Assemblies

Pagina 160 - Closed-Loop Scanners 9

Agilent 5500 SPM Components 2Agilent 5500 SPM User’s Guide 33The major components for the Agilent 5500 SPM are shown in Figure 10.Figure 10 Components

Pagina 161

Agilent 5500 SPM Components 2Agilent 5500 SPM User’s Guide 34MicroscopeThe microscope (Figure 11) includes the hinged support stand, coarse z-axis mot

Pagina 162 - MAC Mode

Agilent 5500 SPM Components 2Agilent 5500 SPM User’s Guide 35ProbesThe SPM techniques described in Chapter 1, “Introduction to the Agilent 5500,” are

Pagina 163 - Connections

Agilent 5500 SPM Components 2Agilent 5500 SPM User’s Guide 36Nose AssemblyThe nose assembly retains the cantilever and enables its motion. A spring cl

Pagina 164 - Hardware and Sample Setup

Agilent 5500 SPM Components 2Agilent 5500 SPM User’s Guide 37damage to the scanner during installation of the nose assembly. Currently the two-piece n

Pagina 165 - MAC Mode 10

Agilent 5500 SPM Components 2Agilent 5500 SPM User’s Guide 38ScannerThe Agilent 5500 is a tip-scanning system, in which the cantilever sits on a scann

Pagina 166 - MAC III Mode

Agilent 5500 SPM Components 2Agilent 5500 SPM User’s Guide 39• The large multi-purpose scanner includes four piezo elements for X and Y and provides s

Pagina 167 - Initial Setup

Read This FirstN9410-90001 Agilent 5500 SPM User’s Guide 4Specifications Environmental ConditionsTemperature (Operating): 5 to 40 °C Temperature (Non-

Pagina 168

Agilent 5500 SPM Components 2Agilent 5500 SPM User’s Guide 40DetectorThe photodiode detector receives the reflection of the laser spot off the back of

Pagina 169 - MAC III Mode 11

Agilent 5500 SPM Components 2Agilent 5500 SPM User’s Guide 41Sample PlatesThe Agilent 5500 SPM is designed to allow scanning from above the sample. A

Pagina 170

Agilent 5500 SPM Components 2Agilent 5500 SPM User’s Guide 42The microscope stand is equipped with three magnetic posts from which a sample plate is m

Pagina 171 - MAC III Software Controls

Agilent 5500 SPM Components 2Agilent 5500 SPM User’s Guide 43Video SystemThe video system lets you locate regions of interest and align the laser on t

Pagina 172 - Contact Mode

Agilent 5500 SPM Components 2Agilent 5500 SPM User’s Guide 44Head Electronics Box (HEB)The Head Electronics Box (HEB) (Figure 20) reads the signals fr

Pagina 173

Agilent 5500 SPM Components 2Agilent 5500 SPM User’s Guide 45AFM ControllerThe AFM Controller (Figure 22) provides the high voltage to the piezoes and

Pagina 174

Agilent 5500 SPM Components 2Agilent 5500 SPM User’s Guide 46Vibration Isolation ChamberThe isolation chamber (Figure 23) isolates the 5500 SPM from v

Pagina 175

Agilent 5500 SPM Components 2Agilent 5500 SPM User’s Guide 47SoftwareThe Agilent 5500 SPM includes PicoView, a powerful software package for controlli

Pagina 176

Agilent 5500 SPM Components 2Agilent 5500 SPM User’s Guide 48System OptionsMany options are available for the Agilent 5500 SPM. As discussed above, pr

Pagina 177

Agilent 5500 SPM Components 2Agilent 5500 SPM User’s Guide 49MAC III ModeMAC III Mode provides the benefits of regular MAC mode, provides three lock-i

Pagina 178

Read This FirstN9410-90001 Agilent 5500 SPM User’s Guide 5logotype (shown below). DO NOT stare directly into the laser beam. To ensure safe operation,

Pagina 179

Agilent 5500 SPM Components 2Agilent 5500 SPM User’s Guide 50Temperature ControlThis option includes low and/or high temperature sample plates, a temp

Pagina 180

Agilent 5500 SPM Components 2Agilent 5500 SPM User’s Guide 51environment, it is possible to perform experiments under very reactive conditions without

Pagina 181

Agilent 5500 SPM Components 2Agilent 5500 SPM User’s Guide 52dynamic properties of biological systems by imaging patterns of molecular binding and adh

Pagina 182

53Agilent 5500 SPMUser’s GuideAgilent Technologies3Setting Up the Agilent 5500 SPMComponent and Facility Dimensions 53Facility Requirements 55Ut

Pagina 183 - Lock-In Tabs

Setting Up the Agilent 5500 SPM 3Agilent 5500 SPM User’s Guide 54• MAC III Controller: 254 mm W x 127 mm H x 254 mm D (10 in W x 5 in H x 10 in D)• Vi

Pagina 184

Setting Up the Agilent 5500 SPM 3Agilent 5500 SPM User’s Guide 55Facility Requirements Following these guidelines for preparing the Agilent 5500 SPM f

Pagina 185 - Outputs Tab

Setting Up the Agilent 5500 SPM 3Agilent 5500 SPM User’s Guide 56Utilities The following table summarizes the utility requirements for the Agilent 550

Pagina 186

Setting Up the Agilent 5500 SPM 3Agilent 5500 SPM User’s Guide 57acoustic noise should be less than 75 dBc (Criterion C). Use of the vibration isolati

Pagina 187

Setting Up the Agilent 5500 SPM 3Agilent 5500 SPM User’s Guide 58Connecting the ComponentsThe cabling for the standard 5500 SPM is shown in Figure 31.

Pagina 188 - Other Tab

59Agilent 5500 SPMUser’s GuideAgilent Technologies4Preparing for ImagingSetting Up the Scanner Assembly 59One-Piece Nose Assembly 60Inserting th

Pagina 189

Read This FirstN9410-90001 Agilent 5500 SPM User’s Guide 6Piezo Scanner Precautions Piezo scanners are, by nature, very FRAGILE pieces of equipment. T

Pagina 190 - Liquid Cell

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 60voltages to opposite piezo elements in the scanner so that one element elongates and the other

Pagina 191 - Liquid Cell 12

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 61scanner mounting fixture (Figure 32). Place the nose assembly in the socket on top of the scann

Pagina 192

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 62Removing the One-Piece Nose AssemblyA removal tool is included with your system to limit damagi

Pagina 193 - Flow-Through Liquid Cell

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 635Once the nose assembly is clear of the scanner you can remove it from the tool.Figure 35 Using

Pagina 194 - Temperature Control

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 64Figure 36 Do not use tweezers to remove a nose assembly. Doing so can place damaging lateral fo

Pagina 195 - Temperature Control 13

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 65Figure 38 Probe properly situated on AFM nose assemblyAFM probe tips are extremely delicate and

Pagina 196

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 66Figure 39 Holding the tweezers as shown, remove a probe from the protective case5 With your fre

Pagina 197

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 67Two-Piece Nose AssemblyThe two-piece nose assembly was designed to simplify the process of inse

Pagina 198

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 68page 60). Place the body in the socket on top of the scanner, aligning its contact pins.Applyin

Pagina 199

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 69Inserting a Probe in the Two-Piece Nose AssemblyAFM probe tips are extremely delicate and can b

Pagina 200 - Do not use double-sided tape

Read This FirstN9410-90001 Agilent 5500 SPM User’s Guide 7Technologies, Inc. Furthermore, Agilent Technologies, Inc. assumes no responsibility or liab

Pagina 201

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 70Figure 44 Move the lever to open the nose assembly disk.4 Place the probe under the copper-colo

Pagina 202

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 71Figure 45 Placing scanner assembly into microscope3 Finger-tighten the knob on the right side o

Pagina 203

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 72Aligning the LaserThe next step is to ensure that the scanner’s laser spot is aligned to reflec

Pagina 204

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 73cantilever tip (counterclockwise) or away from it. The left-to-right knob adjusts the lateral p

Pagina 205

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 74Figure 48 Steps to aligning laser on cantilever beam tip4 Rotate the front-to-back knob counter

Pagina 206 - Water Cooling

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 75The process is similar for triangular-shaped cantilevers, with the exception that the laser wil

Pagina 207

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 76back to the center of its travel in both directions. Doing so should make the laser spot reappe

Pagina 208

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 77Figure 52 shows how the position of the laser on the cantilever affects the position of the las

Pagina 209 - Environmental Control

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 78Note that the IR sensor card should be used for coarse positioning of the laser if using the 98

Pagina 210

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 79Inserting and Aligning the DetectorAs mentioned earlier, the photodiode detector records change

Pagina 211 - Environmental Control 14

Read This FirstN9410-90001 Agilent 5500 SPM User’s Guide 8Declaration of Conformity DECLARATION OF CONFORMITY According to ISO/IEC Guide 22 and CEN/CE

Pagina 212 - Glove Box

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 80microscope base. You can install the detector into the scanner before or after installing the s

Pagina 213 - Electrochemistry

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 81The Friction signal is the difference between the left and right halves divided by the Sum. Fig

Pagina 214 - Modes or STM Mode

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 823For Contact Mode, The dotted yellow line shows the recommended vertical alignment of the laser

Pagina 215 - Electrochemistry 15

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 83Mounting the SampleThe Agilent 5500 SPM accepts a wide variety of sample plates, including spec

Pagina 216 - Equipment

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 84Figure 58 Mounting a sample plate.3 Place the second alignment tab over the alignment pin as in

Pagina 217 - Counter Electrode

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 854Let the magnets on the three posts gently engage the sample plate to hold it in place, as in i

Pagina 218 - Cleaning

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 86Using the Video SystemThe Agilent 5500 SPM includes a USB-based video system for viewing the ti

Pagina 219 - Substrate Cleaning

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 87The z-position knob (Figure 60) lets you raise and lower the video system optics. Typically the

Pagina 220 - Electrochemistry Definitions

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 88clockwise to move the field-of-view to the right, and counterclockwise to move left.Figure 61 A

Pagina 221 - Software Controls

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 89Choose Controls > CameraView to view the video output from the camera (Figure 63).Figure 63

Pagina 222 - Galvanostat

Read This FirstN9410-90001 Agilent 5500 SPM User’s Guide 9Contact Information Agilent Technologies, Inc.4330 W. Chandler Blvd., Chandler, Arizona 8522

Pagina 223 - Appendix A

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 90Care and Handling of the Probes and ScannerProbesAlways store probes at room temperature in the

Pagina 224

Preparing for Imaging 4Agilent 5500 SPM User’s Guide 91to excessive humidity, temperature changes or contact. Agilent recommends that scanners be stor

Pagina 225

92Agilent 5500 SPMUser’s GuideAgilent Technologies5Contact Mode ImagingSetting Up for Contact Mode Imaging 93Constant Force Mode 93Constant Heig

Pagina 226

Contact Mode Imaging 5Agilent 5500 SPM User’s Guide 93Setting Up for Contact Mode ImagingContact Mode imaging can be completed with any of the multi-p

Pagina 227

Contact Mode Imaging 5Agilent 5500 SPM User’s Guide 94Figure 64 Tip in focus through video systemb Lower the focal plane just slightly below the tip b

Pagina 228

Contact Mode Imaging 5Agilent 5500 SPM User’s Guide 9511Locate the area of interest on the sample by manually moving the X/Y stage control micrometers

Pagina 229

Contact Mode Imaging 5Agilent 5500 SPM User’s Guide 96Figure 67 Servo window showing Setpoint voltage and Gains13 Click the Approach button in PicoVie

Pagina 230

Contact Mode Imaging 5Agilent 5500 SPM User’s Guide 97changes in tip deflection in order to maintain constant force. 10 % is a good starting value; mo

Pagina 231

Contact Mode Imaging 5Agilent 5500 SPM User’s Guide 98buffer frame, then select the Input Signal from the list (Figure 69). The list will vary dependi

Pagina 232

Contact Mode Imaging 5Agilent 5500 SPM User’s Guide 99Figure 70 Scan and Motor window after scan has been initiatedAs the tip moves across the first s