Agilent Technologies E1445A Guía de usuario Pagina 283

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 539
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 282
Agilent E6432A Microwave Synthesizer Users Guide 4-47
stepFreqHz
Variable Type ViReal64
This parameter specifies the calibration step size for the external
leveling loop configuration being calibrated. Its value is used in
conjunction with the start and stop frequency values that are also
specified.
The calibration step size is used to determine the actual points used in
the External Modulator Gain Calibration. The smaller the step size, the
more accurate the calibration is for all frequencies, but this enhanced
accuracy is at the cost of increased time for the calibration process. A
calibration step size value of 100 MHz is recommended.
The number of calibration steps is defined by the following equation
and can not exceed 201:
NOTE If other combinations of hardware are being used in the external
leveling loop configuration that do not have an operational range equal
to the external detector being used, their operational range must be
used instead.
Return Value
This return value reports the status of the Generate/Load Ext Freq
Table function.
To obtain further information about the status that is returned, call the
HPE6432_error_message function or refer to “Error-Code and
Fail-Code Messages” on page 3-93.
Related Topics
HPE6432_error_message
Error-Code and Fail-Code Messages
Reset Ext Det Cal Data
Power Search
Specifications and Characteristics
N
umber of Steps Stop Frequency Start Frequency()Step Frequency=
Vista de pagina 282
1 2 ... 278 279 280 281 282 283 284 285 286 287 288 ... 538 539

Comentarios a estos manuales

Sin comentarios