Agilent Technologies E4438C Guía de instalación Pagina 35

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 56
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 34
Operation Verification
AM Accuracy Check
Chapter 3 31
AM Accuracy Check
Connect the Test Equipment
Figure 3-5 AM Accuracy Equipment Setup
Configure the Signal Generator
1. Press Preset.
2. Press
AM > AM Off On. The AM annunciator is displayed.
3. Press
AM Rate. Enter 1 using the numeric keypad and press the kHz terminator softkey.
4. Press
Amplitude. Enter 7 using the numeric keypad and press the dBm terminator softkey.
5. Ensure that the MOD ON annunciator is displayed. If it is not, press
Mod On/Off.
6. Press
RF On/Off. The RF ON annunciator is displayed.
Configure the Measuring Receiver
1. Reset the measuring receiver.
2. Set the measuring receiver to AM mode.
3. Turn on Peak+ detector.
4. Turn on the 300 Hz high-pass filter.
5. Turn on the 3 kHz low-pass filter.
Vista de pagina 34
1 2 ... 30 31 32 33 34 35 36 37 38 39 40 ... 55 56

Comentarios a estos manuales

Sin comentarios