Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Firmware Upgrade Guide
Step 3. Connect the Signal Generator to the PC
6
Step 3. Connect the Signal Generator to the PC
NOTE The N5161A/62A MXG ATE is identical to the N5181A/82A with the exception that they do not have front
panel functionality (no display or keys). Instead all functionality is controlled via SCPI commands or through
the Web- Enabled MXG. For signal generator functionality, refer to the Users Guide. For information on
Web–Enabled MXG, refer to the Installation Guide, the Programming Guide and the SCPI Command
Reference.
Using LAN
Connect the signal generator to the PC:
Either
use a crossover cable to connect the signal generator directly to the PC,
or
connect the PC to the LAN and use a 100Base–T LAN cable to connect the signal generator to the DHCP Server.
There are numerous ways to configure the signal generator to connect to the LAN. Refer to:
Agilent Technologies IO Libraries USB/LAN/GPIB Interfaces Connectivity Guide (E2094–90009).
Agilent MXG FAQ’s “How do I connect to the LAN?”.
Agilent MXG Programming Guide or,
www.agilent.com/find/mxg.
Depending on the signal generator's current LAN settings and which procedure you choose, the signal generator's LAN
settings may be changed. Since these settings are persistent (retained through presets and power cycles) it is
recommended that you record the current LAN settings so you can restore them back to their original settings after
upgrading the firmware. Information on the LAN softkeys is provided in key
Help of the instrument. See also the
User’s Guide.
Using GPIB
1. Connect a GPIB cable between the PC and the signal generator.
2. Set the GPIB address:
Press
Utility > I/O Config > GPIB Setup > GPIB Address > 19
The signal generator’s GPIB address is set to 19 at the factory. The acceptable range of addresses is 0 through 30.
Once initialized, the state of the GPIB address is not affected by a signal generator preset or by a power cycle.
Other instruments on the GPIB cannot use the same address as the signal generator.
NOTE If there is more than one signal generator on the same bus whose GPIB address is 19, turn the power off to
the instrument(s) or change the address of the signal generator of interest.
Using USB
Connect a USB cable between the PC and the signal generator’s rear panel mini–B Devise USB connector.
If the “Found New Hardware” wizard activates, select
Install the Software Automatically.