Agilent Technologies DisplayPort Manual de servicio Pagina 113

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 246
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 112
Source Spread Spectrum Clocking (SSC) Differential Tests (Normative & Informative) 13
Agilent U7232A DisplayPort Electrical Performance Compliance Test Application 113
.
You can use any oscilloscope channel and connect it to any lane test point.
You select the channels used for testing lanes in the Set Up tab of the
DisplayPort Electrical Performance Compliance Test Application.
The data lanes and channels shown in the previous figures are just
examples. You can choose any desired data lane and channel that you
want.
For more information on the 1168A or 1169A probe amplifiers and
differential probe heads, see Chapter 25, “InfiniiMax Probing,” starting on
page 231.
Infiniium Oscilloscope
Data Channel
SMA to SMP cables
DUT
Lane on the W2641A test fixture
Figure 38 Differential Measurement Setup Using Two Single Ended Connections - SSC
Tests (A Minus B Configuration)