Source Unit Interval Differential Tests (Informative) 11
Agilent U7232A DisplayPort Electrical Performance Compliance Test Application 97
.
You can use any oscilloscope channel and connect it to any lane test point.
You select the channels used for testing lanes in the Set Up tab of the
DisplayPort Electrical Performance Compliance Test Application.
The data lanes and channels shown in the previous figures are just
examples. You can choose any desired data lane and channel that you
want.
Infiniium Oscilloscope
Data Channel
SMA to SMP cables
DUT
Lane on the W2641A test fixture
Figure 32 Differential Measurement Setup Using Two Single Ended Connections - Unit
Interval Tests (A Minus B Configuration)