Calibration of the Touch Screen
2. Installation
Calibration of the Touch Screen
After E5071C measurement screen appears, you have to calibrate the touch screen. Follow
the procedure described below to calibrate the touch screen.
Step 1. Press .
Step 2. Press Service Menu.
Step 3. Press Test Menu.
Step 4. Press Adjust Touch Screen.
The touch screen calibration screen (Figure 2-10) appears.
Figure 2-10 Touch Panel Calibration Screen
e5071cse1078
Step 5. Touch the x mark on the upper left with your finger. The mark x appears also on the upper
right, lower left, and lower right. Touch the x marks in that order with your finger.
Touching the four locations described above with your finger automatically concludes the
touch screen calibration.
NOTE With no operation on the touch screen calibration screen for a preset time, it automatically
closes and the previous measurement screen reappears.