7
Agilent N4916B Specifications
General characteristics
Operating temperature 5 ºC to 40ºC (–23 ºF to 104 ºF)
Storage temperature –40 ºC to +70ºC (–65 ºF to 158 ºF)
Operating Humidity 95% relative humidity, non-condensing
Storage humidity 50 % relative humidity
Power requirements 100 V to 240 V, ±10 %, 47 Hz to 63 Hz, 80 VA
Physical dimensions WxHxD Bench top (with bumper) 228 x 59 x 246 mm
Rack mount (without bumper) ½* 19” width, 1U height: 213 x 44.5 x 245 mm
Weight net 2.05 kg
Weight shipping
tbd
Recommended recalibration period 1 year recommended
Warranty period 1 year return to Agilent. See ordering instructions for extended warranty.
Regulatory standards
Safety IEC61010-1:2001, EN61010-1:2001, CAN/CSA-C22 No. 61010-04, UL 61010-1:2004
EMC IEC61326-1:1997+A1:1998, EN61326-1:1997+A1:1998
Quality management ISO 9004, ISO 14000
Environmental ROHS compliant
Remote control interfaces
Connectivity USB 2.0, rear panel, LAN
Programming language SCPI
Via J-BERT Via USB2 to the controlling J-BERT N4903B, which provides LAN, GPIB, USB as remote control
interfaces. The N4903B requires software revision 6.5x or higher to control the N4916B.
Stand-alone user interface
System requirements OS: Microsoft Windows
®
(verified on XP, SP2), Agilent I/O Libraries Suite rev.15.5, Microsoft .NET 2.0
Specification assumptions
The specifications in this document describe the instruments warranted performance. Preliminary values are shown in italic.
Non-warranted values are described as typical. All specifications are valid in a range from 5 ºC to 40 ºC ambient temperature
after a warm-up time of 30 minutes. If not otherwise stated, all inputs and outputs need to be terminated with 50 Ω to GND.
Ordering information
4-tap de-emphasis signal converter N4916B
Clock multiplier N4916B-001
Matched cable pair for connecting data and
clock input to J-BERT N4903B
N4915A-010
Adapter 3.5 mm (f) to 2.4 mm (m) (1 each
is needed for connecting N4916B with
N4915A-010 cable kit to N4903A or ParBERT)
N4911A-002
Rack mount kit E5810A-100
Warranty and calibration services R1280, R1282
Productivity assistance R1380-N49xx
Accessories included in N4916B 4 x 50 Ω terminations 3.5 mm, USB cable, Test report “UK6”, USB-stick with
software and user documentation
Comentarios a estos manuales