3 Programming the Analyzer
IEEE 488.2 Common Commands
Remote Command
*TRG
Example *TRG Triggers the instrument to take a sweep or start a measurement, depending on the current
instrument settings.
Notes See related command :INITiate:IMMediate.
Initial S/W Revision Prior to A.02.00
Self Test Query
This query performs the internal self-test routines and returns a number indicating the success of
the testing. A zero is returned if the test is successful, 1 if it fails.
Remote Command
*TST?
Example *TST? Runs the self-test routines and returns 0=passed, 1=some part failed.
Initial S/W Revision Prior to A.02.00
Wait-to-Continue
This command causes the instrument to wait until all overlapped commands are completed before
executing any additional commands. There is no query form for the command.
Remote Command
*WAI
Example INIT:CONT OFF; INIT;*WAI Sets the instrument to single sweep. Starts a sweep and waits for its
completion.
Status Bits/OPC
dependencies
Not global to all remote ports or front panel. *OPC only considers operation that was initiated on
the same port as the *OPC command was issued from.
Initial S/W Revision Prior to A.02.00
110 Remote Language Compatibility Measurement Application Reference