3 Programming the Analyzer
STATus Subsystem
"Questionable Negative Transition " on page 145
"Questionable Positive Transition" on page 146
Questionable Condition
This query returns the decimal value of the sum of the bits in the Questionable Condition register.
The data in this register is continuously updated and reflects the current conditions.
Mode All
Remote Command
:STATus:QUEStionable:CONDition?
Example STAT:QUES:COND?
Preset 0
Status Bits/OPC dependencies Sequential command
Initial S/W Revision Prior to A.02.00
Questionable Enable
This command determines which bits in the Questionable Event register will set the Questionable
Status Summary bit (bit3) in the Status Byte Register. The variable <integer> is the sum of the
decimal values of the bits you want to enable.
The preset condition is all bits in this enable register set to 0. To have any Questionable Events reported to
the Status Byte Register, one or more bits need to be set to 1. The Status Byte Event Register should be
queried after each measurement to check the Questionable Status Summary (bit 3). If it is equal to 1, a
condition during the test may have made the test results invalid. If it is equal to 0, this indicates that no
hardware problem or measurement problem was detected by the analyzer.
Mode All
Remote Command
:STATus:QUEStionable:ENABle 16
:STATus:QUEStionable:ENABle?
Example STAT:OPER:PTR 1 Align Now operation beginning will be reported to the Status Byte Register.
Preset 0
Min 0
Max 32767
Status Bits/OPC Sequential command
144 Remote Language Compatibility Measurement Application Reference