
3 Programming the Analyzer
STATus Subsystem
Questionable Positive Transition
This command determines which bits in the Questionable Condition register will set the
corresponding bit in the Questionable Event register when the condition register bit has a positive
transition (0 to 1). The variable <integer> is the sum of the decimal values of the bits that you
want to enable.
Mode All
Remote Command
:STATus:QUEStionable:PTRansition <integer>
:STATus:QUEStionable:PTRansition?
Example STAT:QUES:PTR 16 Temperature summary ‘questionable asserted’ will be reported to the Status
Byte Register.
Preset 32767
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision Prior to A.02.00
Questionable Calibration Register
The following commands and queries are available for this register:
"Questionable Calibration Condition" on page 146
"Questionable Calibration Enable" on page 147
"Questionable Calibration Event Query" on page 147
"Questionable Calibration Negative Transition" on page 148
"Questionable Calibration Positive Transition" on page 148
Questionable Calibration Condition
This query returns the decimal value of the sum of the bits in the Questionable Calibration
Condition register.
The data in this register is continuously updated and reflects the current conditions.
146 Remote Language Compatibility Measurement Application Reference
Comentarios a estos manuales