Hub Hi-Speed Tests 4
Notes on USB Electrical Compliance Testing 125
2 Click OK to close the Test Instructions dialog.
EL_28 Hub CHIRP Response to Reset from Suspend
The USB automated test application will prompt you to perform these
steps:
1 On the Hub Test menu of the HS Electrical Test Tool software, select
SUSPEND from the Hub Command drop down menu. Click [EXECUTE]
once to place the device into suspend.
2 Click OK to close the Test Instructions dialog.