Agilent Technologies N5416A Manual de servicio Pagina 55

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 236
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 54
Device Hi-Speed Tests 3
Notes on USB Electrical Compliance Testing 55
2 Click OK to close the Test Instructions dialog.
EL_27 Device CHIRP Response to Reset from Hi-Speed Operation
The USB automated test application will prompt you to perform these
steps:
1 On the Device Test Menu of the HS Electrical Test Tool, select RESET
from the Device Command drop down menu. Click [EXECUTE] once to
reset the device operating in high speed. The captured transition should
be as in the figure below.
Vista de pagina 54
1 2 ... 50 51 52 53 54 55 56 57 58 59 60 ... 235 236

Comentarios a estos manuales

Sin comentarios