Agilent Technologies N5416A Manual de servicio Pagina 166

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 236
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 165
166 Notes on USB Electrical Compliance Testing
5 Host Hi-Speed Electrical Tests
9 Check I have completed these instructions.
Running the Tests
1 Click Run Tests.
Test Instructions
EL_39 Suspend Timing Response
The USB automated test application will prompt you to perform these
steps:
1 On the Host Test menu of the HS Electrical Test Tool software, click
[Enumerate Bus] once.
2 Select SUSPEND from the Port Control drop down menu.
3 Enter the port number. Click [EXECUTE] once to place the port into
suspend. The captured transition should be as in the figure below.
Vista de pagina 165
1 2 ... 161 162 163 164 165 166 167 168 169 170 171 ... 235 236

Comentarios a estos manuales

Sin comentarios