102 16800 Series Portable Logic Analyzers Service Guide
5 Troubleshooting
Configure the logic analyzer to test Pod 1
1 Exit the logic analysis application (from the main menu,
choose FileExit) and then restart the application. This
puts the logic analyzer into its initial state.
2 Set up the bus and signals to test Pod 1.
a From the Agilent Logic Analyzer application’s main
menu, select SetupBus/Signal... or SetupMy Logic
Analyzer-1Bus/Signal....
b In the Analyzer Setup dialog, ensure that the Threshold
button for Pod 1 is set to TTL (1.50 V).
c Verify that the activity indicators (the red arrows)
show activity on all 16 channels that are connected to
the pulse generator.
d Assign all channels. Hint: you can do this quickly by
clicking on the left- most channel, then dragging to the
right across all of the other channels. If you have a
touch screen model (Option 103), you can touch the
touchscreen and drag across with your finger.
Threshold