Agilent Technologies 8133A Manual de servicio Pagina 116

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 188
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 115
116 16800 Series Portable Logic Analyzers Service Guide
6 Replacing Assemblies
To prepare the instrument for disassembly
Do this procedure before doing any disassembly procedure on
the instrument.
1 Close the Agilent Logic Analyzer application.
2 Power off the system using one of the methods explained
below.
3 Remove the power cord.
4 Move the instrument to a static safe work environment.
To power off the system
There are several ways to power off the logic analyzer.
Using the Windows Shutdown
On the logic analyzer desktop click Start>Shut Down or if you
are running remote desktop, click Start>Settings>Windows
Security>Shut Down. This software power off does the
following:
Closes all programs that are running.
Writes all data to the disk.
Turns off the power supply.
WARNING
Hazardous voltages exist on the power supply. To avoid electrical
shock, disconnect power from the instrument before performing
the following procedures. After disconnecting the power, wait at
least six minutes for the capacitors on the power supply board to
discharge before servicing the instrument.
CAUTION
Damage can occur to electronic components if you remove or
replace assemblies when the instrument is on or when the power
cable is connected. Never attempt to remove or install any assembly
with the instrument on or with the power cable connected.
NOTE
When powering off the logic analyzer, wait until the fans stop turning
(about 15 seconds) before turning the logic analyzer back on. This
ensures that internal circuitry restarts in a known state.
Vista de pagina 115
1 2 ... 111 112 113 114 115 116 117 118 119 120 121 ... 187 188

Comentarios a estos manuales

Sin comentarios