48 16800 Series Portable Logic Analyzers Service Guide
3 Testing Performance
g Ensure that the Clock Mode is set to Master.
h Set the Pod 1 master clock to Both Edges.
Adjust the sample positions using eye finder
1 Click Thresholds and Sample Positions....
2 In the “Buses/Signals” section of the Thresholds and
Sample Positions dialog, ensure that the check box next to
“My Bus 1” is checked.
NOTE
If option 500 is not installed on the logic analyzer, then 250 MHz will be the
only speed available.