46 16800 Series Portable Logic Analyzers Service Guide
3 Testing Performance
c In the Threshold Settings dialog, set the threshold
value for Pod 1 of the logic analyzer to 1 V, and click
OK.
d The activity indicators now show activity on the
channels that are connected to the pulse generator.
e Un- assign all channels. Hint: you can do this quickly by
clicking on the left- most check mark and dragging to
the right across all of the other check marks. If you
have a logic analyzer with a touchscreen (Option 103),
you can touch the touchscreen and drag across with
your finger.
f Click (or touch) to select channels 2, 6, 10 and 14 as
shown.
g Drag the scroll bar all the way to the left and ensure
that the activity indicator shows activity on clock 1.